Subject motion may cause errors in estimates of blood T when using the T -relaxation under spin tagging (TRUST) technique on noncompliant subjects like neonates. By incorporating 3D volume navigators (vNavs) into the TRUST pulse sequence, independent measurements of motion during scanning permit evaluation of these errors.
The effects of integrated vNavs on TRUST-based T estimates were evaluated using simulations and in vivo subject data. Two subjects were scanned with the TRUST+vNav sequence during prescribed movements. Mean motion scores were derived from vNavs and TRUST images, along with a metric of exponential fit quality. Regression analysis was performed between T estimates and mean motion scores. Also, motion scores were determined from independent neonatal scans.
vNavs negligibly affected venous blood T estimates and better detected subject motion than fit quality metrics. Regression analysis showed that T is biased upward by 4.1 ms per 1 mm of mean motion score. During neonatal scans, mean motion scores of 0.6 to 2.0 mm were detected.
Motion during TRUST causes an overestimate of T , which suggests a cautious approach when comparing TRUST-based cerebral oxygenation measurements of noncompliant subjects. Magn Reson Med 78:2283-2289, 2017. © 2017 International Society for Magnetic Resonance in Medicine.
在对不配合的受试者(如新生儿)使用自旋标记(TRUST)技术下的 T -弛豫时间(T )估计时,由于受试者运动,可能会导致 T 估计值出现误差。通过将 3D 容积导航(vNavs)纳入 TRUST 脉冲序列,在扫描过程中可以独立测量运动,从而评估这些误差。
使用模拟和体内受试者数据评估了集成 vNavs 对基于 TRUST 的 T 估计的影响。使用 TRUST+vNav 序列对 2 名受试者进行扫描,同时进行规定的运动。从 vNavs 和 TRUST 图像中得出平均运动评分,以及指数拟合质量的度量标准。对 T 估计值和平均运动评分之间进行回归分析。此外,还从独立的新生儿扫描中确定了运动评分。
vNavs 对静脉血 T 估计值的影响可以忽略不计,且比拟合质量度量标准更好地检测到受试者运动。回归分析表明,T 每增加 1 毫米的平均运动评分,就会向上偏倚 4.1 毫秒。在新生儿扫描期间,检测到 0.6 至 2.0 毫米的平均运动评分。
TRUST 过程中的运动导致 T 的高估,这表明在比较不配合的受试者的基于 TRUST 的脑氧合测量值时,应谨慎使用。磁共振医学 78:2283-2289, 2017。© 2017 国际磁共振学会。