单一设计的大头模块化金属对金属髋关节置换的检索分析。

Retrieval Analysis of Large-Head Modular Metal-on-Metal Hip Replacements of a Single Design.

机构信息

Orthopaedic Innovation Centre, Winnipeg, Manitoba, Canada.

Concordia Joint Replacement Group, Winnipeg, Manitoba, Canada; Department of Surgery, University of Manitoba, Winnipeg, Manitoba, Canada.

出版信息

J Arthroplasty. 2018 Jun;33(6):1945-1952. doi: 10.1016/j.arth.2017.12.044. Epub 2018 Jan 9.

Abstract

BACKGROUND

There are limited publications examining modular metal-on-metal (MoM) total hip implants in which a comprehensive analysis of retrieved components is performed. This study examines 24 retrieved modular MoM implants from a single manufacturer and compares retrieval analytics; bearing surface damage, wear, and modular taper corrosion against patient, surgical and implant characteristics to elucidate significant associations.

METHODS

Clinical, patient, and surgical data were collected including age, body mass index, blood metal ion levels, and cup inclination. Damage assessment was performed visually in addition to surface profilometry. Acetabular liners and femoral heads were measured for volumetric wear. Femoral head taper bores were similarly measured for material removal due to corrosion and fretting.

RESULTS

Patients with MoM-related reasons for revision showed significantly higher levels of blood metal ion levels. Bearing wear was strongly associated with blood metal ion levels and was significantly increased in cups placed more vertically. Younger patients tended to have higher body mass indices as well as poorer cup placement.

CONCLUSION

This work details a broad range of analyses on a series of modular MoM total hip implants from a single manufacturer of which there are few published studies. Acetabular cup inclination angle was deemed a primary cause of revision surgery through increased MoM wear, high metal ion levels in the blood, and subsequent adverse local tissue reactions. Heavy patients can increase the surgical difficulty which was shown to be related to poor cup placement in this cohort.

摘要

背景

目前关于模块化金属对金属(MoM)全髋关节植入物的研究有限,这些研究通常只对所取出的部件进行综合分析。本研究对来自同一家制造商的 24 个取出的模块化 MoM 植入物进行了检查,并对取出的分析物;对轴承表面损伤、磨损和模块化锥度腐蚀与患者、手术和植入物特征进行了对比分析,以阐明显著关联。

方法

收集了临床、患者和手术数据,包括年龄、体重指数、血液金属离子水平和杯倾斜度。除了表面轮廓测量外,还进行了视觉损伤评估。对髋臼衬垫和股骨头进行了体积磨损测量。同样,对股骨头锥度孔进行了由于腐蚀和微动导致的材料去除测量。

结果

因 MoM 相关原因进行翻修的患者血液金属离子水平明显更高。轴承磨损与血液金属离子水平密切相关,在更垂直放置的杯子中磨损明显增加。年轻患者的体重指数往往更高,杯的位置也更差。

结论

这项工作详细介绍了对来自同一家制造商的一系列模块化 MoM 全髋关节植入物进行的广泛分析,目前很少有关于此类研究的出版物。通过增加 MoM 磨损、血液中金属离子水平升高以及随后的局部组织不良反应,认为髋臼杯倾斜角度是导致翻修手术的主要原因。肥胖患者会增加手术难度,这在本研究中表现为杯的位置不佳。