未经医学培训观察者的视觉听诊法血压测量准确性和可变性的定量评估。

Quantitative Assessment of Blood Pressure Measurement Accuracy and Variability from Visual Auscultation Method by Observers without Receiving Medical Training.

机构信息

Department of Cardiology, Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, China.

Department of Electrical and Electronic Engineering, Southern University of Science and Technology, Shenzhen, China.

出版信息

Biomed Res Int. 2017;2017:3537079. doi: 10.1155/2017/3537079. Epub 2017 Dec 20.

Abstract

This study aimed to quantify blood pressure (BP) measurement accuracy and variability with different techniques. Thirty video clips of BP recordings from the BHS training database were converted to Korotkoff sound waveforms. Ten observers without receiving medical training were asked to determine BPs using (a) traditional manual auscultatory method and (b) visual auscultation method by visualizing the Korotkoff sound waveform, which was repeated three times on different days. The measurement error was calculated against the reference answers, and the measurement variability was calculated from the SD of the three repeats. Statistical analysis showed that, in comparison with the auscultatory method, visual method significantly reduced overall variability from 2.2 to 1.1 mmHg for SBP and from 1.9 to 0.9 mmHg for DBP (both < 0.001). It also showed that BP measurement errors were significant for both techniques (all < 0.01, except DBP from the traditional method). Although significant, the overall mean errors were small (-1.5 and -1.2 mmHg for SBP and -0.7 and 2.6 mmHg for DBP, resp., from the traditional auscultatory and visual auscultation methods). In conclusion, the visual auscultation method had the ability to achieve an acceptable degree of BP measurement accuracy, with smaller variability in comparison with the traditional auscultatory method.

摘要

本研究旨在定量评估不同技术测量血压(BP)的准确性和变异性。将 BHS 培训数据库中的 30 段 BP 记录视频片段转换为柯氏音波形。10 名未经医学培训的观察者被要求使用(a)传统手动听诊法和(b)通过可视化柯氏音波形进行的视觉听诊法来确定 BP,这两种方法在不同的日子里重复了三次。根据参考答案计算测量误差,从三次重复的 SD 计算测量变异性。统计分析表明,与听诊法相比,视觉法显著降低了收缩压的整体变异性,从 2.2mmHg 降至 1.1mmHg(均<0.001),舒张压的整体变异性也从 1.9mmHg 降至 0.9mmHg(均<0.001)。它还表明,两种技术的 BP 测量误差均显著(除传统方法的舒张压外,均<0.01)。虽然显著,但总体平均误差较小(传统听诊法和视觉听诊法的收缩压分别为-1.5mmHg 和-1.2mmHg,舒张压分别为-0.7mmHg 和 2.6mmHg)。总之,与传统听诊法相比,视觉听诊法具有实现可接受的 BP 测量准确性的能力,并且变异性更小。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8cab/5750510/28b78ea8aa16/BMRI2017-3537079.001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验