Ditommaso Savina, Giacomuzzi Monica, Ricciardi Elisa, Garbuio Roberto, Zotti Carla M
Department of Public Health and Pediatrics, University of Turin, Piazza Polonia 94, 10126 Turin, Italy E-mail:
Freelance Dentist, Piazza Campidoglio 2, 10036 Settimo Torinese, Italy.
J Water Health. 2018 Feb;16(1):150-158. doi: 10.2166/wh.2017.035.
In this study we evaluated (1) the efficacy of a protocol that combines hydrogen peroxide (shock treatment) and ICX tablets (continuous treatment) for the control of microbial contamination in dental unit water lines, and (2) the in vitro antimicrobial activity of ICX tablets on collection and wild strains isolated from dental chair output waters. To assess the treatment effectiveness, the microbial load in the output water samples of three dental chairs were investigated: one control chair received only shock treatment. In vitro bactericidal activity was tested against Staphylococcus aureus and Pseudomonas aeruginosa. Data obtained from samples collected from chairs treated with ICX and shock treatment and data from the control chair did not differ significantly on the basis of microbial load. In the in vitro study, the product was unable to kill Gram-negative bacteria. These results show that the continuous introduction of ICX was not effective in maintaining low counts of the heterotrophic bacteria in the output water of dental devices, and shock treatment may be needed more frequently than monthly.
在本研究中,我们评估了:(1)一种将过氧化氢(冲击处理)和ICX片剂(持续处理)相结合的方案对控制牙科设备水线中微生物污染的效果,以及(2)ICX片剂对从牙科椅出水分离出的采集菌株和野生菌株的体外抗菌活性。为评估治疗效果,我们调查了三把牙科椅出水样本中的微生物负荷:一把对照椅仅接受冲击处理。针对金黄色葡萄球菌和铜绿假单胞菌测试了体外杀菌活性。从接受ICX和冲击处理的椅子采集的样本数据与对照椅的数据在微生物负荷方面无显著差异。在体外研究中,该产品无法杀灭革兰氏阴性菌。这些结果表明,持续引入ICX对于维持牙科设备出水中低数量的异养菌无效,且可能需要比每月更频繁地进行冲击处理。