Sun Jun, Liu Yan-Cheng, Yan Song-Hua, Wang Sha-Sha, Lester D Kevin, Zeng Ji-Zhou, Miao Jun, Zhang Kuan
Department of School of Biomedical Engineering, Capital Medical University, Beijing, China.
Beijing Key Laboratory of Fundamental Research on Biomechanics in Clinical Application, Capital Medical University, Beijing, China.
Orthop Surg. 2018 Feb;10(1):32-39. doi: 10.1111/os.12367. Epub 2018 Feb 12.
The third generation Intelligent Device for Energy Expenditure and Activity (IDEEA3, MiniSun, CA) has been developed for clinical gait evaluation, and this study was designed to evaluate the accuracy and reliability of IDEEA3 for the gait measurement of lumbar spinal stenosis (LSS) patients.
Twelve healthy volunteers were recruited to compare gait cycle, cadence, step length, velocity, and number of steps between a motion analysis system and a high-speed video camera. Twenty hospitalized LSS patients were recruited for the comparison of the five parameters between the IDEEA3 and GoPro camera. Paired t-test, intraclass correlation coefficient, concordance correlation coefficient, and Bland-Altman plots were used for the data analysis.
The ratios of GoPro camera results to motion analysis system results, and the ratios of IDEEA3 results to GoPro camera results were all around 1.00. All P-values of paired t-tests for gait cycle, cadence, step length, and velocity were greater than 0.05, while all the ICC and CCC results were above 0.950 with P < 0.001.
The measurements for gait cycle, cadence, step length, velocity, and number of steps with the GoPro camera are highly consistent with the measurements with the motion analysis system. The measurements for IDEEA3 are consistent with those for the GoPro camera. IDEEA3 can be effectively used in the gait measurement of LSS patients.
第三代能量消耗与活动智能设备(IDEEA3,MiniSun,加利福尼亚州)已被开发用于临床步态评估,本研究旨在评估IDEEA3对腰椎管狭窄症(LSS)患者步态测量的准确性和可靠性。
招募12名健康志愿者,比较运动分析系统和高速摄像机之间的步态周期、步频、步长、速度和步数。招募20名住院LSS患者,比较IDEEA3和GoPro摄像机之间的这五个参数。采用配对t检验、组内相关系数、一致性相关系数和Bland-Altman图进行数据分析。
GoPro摄像机结果与运动分析系统结果的比率,以及IDEEA3结果与GoPro摄像机结果的比率均在1.00左右。步态周期、步频、步长和速度的配对t检验的所有P值均大于0.05,而所有ICC和CCC结果均高于0.950,P<0.001。
GoPro摄像机对步态周期、步频、步长、速度和步数的测量与运动分析系统的测量高度一致。IDEEA3的测量结果与GoPro摄像机的测量结果一致。IDEEA3可有效用于LSS患者的步态测量。