Abstract

BACKGROUND

High-flow nasal cannula (HFNC) oxygen therapy produces noise at a level such that patients often complain. However, the noise level has not been measured digitally.

METHODS

We evaluated 3 types of HFNCs without filters and 2 types with filters attached for noise reduction. Optiflow (with and without a filter), MaxVenturi (with and without a filter) and AIRVO2 (without a filter only) were positioned at the center of a hospital room. We measured the noise levels at the distance of 1 m from the equipment at various total flows (30, 40, 50, 60 L/min) and F (0.40, 0.60, and 0.90).

RESULTS

Noise levels were increased with the AIRVO2 and MaxVenturi when total flow and F were increased. Noise levels decreased with the MaxVenturi when a filter was used. The noise level did not change with the Optiflow when total flow and F were increased. The noise level decreased in the groups with AIRVO2 and Optiflow compared with MaxVenturi without a filter.

CONCLUSIONS

The findings in this study show that the noise level of HFNC/Venturi could be reduced by attaching an intake filter. However, the noise level of HFNC/blender and HFNC/turbine decreased in comparison with HFNC/Venturi without an intake filter.

摘要

背景

高流量鼻导管(HFNC)氧疗会产生一定程度的噪音,患者常因此抱怨。然而,噪音水平尚未进行数字测量。

方法

我们评估了3种不带过滤器的HFNC和2种带降噪过滤器的HFNC。将Optiflow(带和不带过滤器)、MaxVenturi(带和不带过滤器)以及AIRVO2(仅不带过滤器)放置在医院病房中央。我们在不同总流量(30、40、50、60升/分钟)和F(0.40、0.60和0.90)条件下,测量了距离设备1米处的噪音水平。

结果

当总流量和F增加时,AIRVO2和MaxVenturi的噪音水平升高。使用过滤器时,MaxVenturi的噪音水平降低。当总流量和F增加时,Optiflow的噪音水平不变。与不带过滤器的MaxVenturi相比,AIRVO2和Optiflow组的噪音水平降低。

结论

本研究结果表明,通过安装进气过滤器可降低HFNC/文丘里管的噪音水平。然而,与不带进气过滤器的HFNC/文丘里管相比,HFNC/混合器和HFNC/涡轮机的噪音水平降低。