Pettigrew R I, Dannels W, Galloway J R, Pearson T, Millikan W, Henderson J M, Peterson J, Bernardino M E
AJR Am J Roentgenol. 1987 Feb;148(2):411-4. doi: 10.2214/ajr.148.2.411.
For evaluation of the feasibility and clinical potential of using the phase data from standard MR imaging sequences to measure blood flow, 11 vessels with diameters of 4 to 7 mm were imaged in seven dogs. The flow in either the superior mesenteric vein or the inferior vena cava was measured first at laparotomy (in ml/min) with electromagnetic flow meters. Immediately thereafter, these vessels were imaged by MR in 25-mm thick sections by using a standard spin echo (SE) 750/30 sequence with a Philips 0.5-T imager. Previous phase-flow calibration of the imager and sequence allowed calculation of the blood flow rates from the phase images that were used to measure the vessels' cross-sectional areas and blood phase values. Comparison of the measurements obtained with each technique showed a significant correlation (r = .977, p less than .05) between MR-imaging values and flow-meter measurements when the blood velocity was less than approximately 40 cm/sec, the known upper limit of the flow dynamic range for the MR hardware and sequence used. There was no correlation for blood velocities greater than 40 cm/sec. However, the range of blood flow velocities in dogs and man extends to more than 100 cm/sec. Thus, these results suggest that this technique might yield valuable adjunctive flow data in routine clinical imaging provided that improvements in hardware and software permit a larger dynamic range.
为评估利用标准磁共振成像序列的相位数据测量血流的可行性和临床潜力,在7只犬身上对11条直径为4至7毫米的血管进行了成像。首先在剖腹手术时用电磁流量计测量肠系膜上静脉或下腔静脉的血流(以毫升/分钟为单位)。此后立即使用飞利浦0.5-T成像仪,通过标准自旋回波(SE)750/30序列,以25毫米厚的层面,对这些血管进行磁共振成像。成像仪和序列先前的相位-血流校准使得能够从用于测量血管横截面积和血相位值的相位图像中计算出血流率。当血流速度小于约40厘米/秒时(这是所用磁共振硬件和序列血流动力学范围的已知上限),对两种技术获得的测量结果进行比较显示,磁共振成像值与流量计测量值之间存在显著相关性(r = 0.977,p < 0.05)。对于血流速度大于40厘米/秒的情况则没有相关性。然而,犬和人的血流速度范围可延伸至超过100厘米/秒。因此,这些结果表明,只要硬件和软件有所改进,允许更大的动态范围,这项技术可能会在常规临床成像中提供有价值的辅助血流数据。