Extending neutron autoradiography technique for boron concentration measurements in hard tissues.

作者信息

Provenzano Lucas, Olivera María Silvina, Saint Martin Gisela, Rodríguez Luis Miguel, Fregenal Daniel, Thorp Silvia I, Pozzi Emiliano C C, Curotto Paula, Postuma Ian, Altieri Saverio, González Sara J, Bortolussi Silva, Portu Agustina

机构信息

Comisión Nacional de Energía Atómica (CNEA), Av. Del Libertador 8250, C1429BNP CABA, Argentina; Consejo Nacional de Investigaciones Científicas y Técnicas (CONICET), Godoy Cruz 2290, C1425FQB CABA, Argentina.

Comisión Nacional de Energía Atómica (CNEA), Av. Del Libertador 8250, C1429BNP CABA, Argentina.

出版信息

Appl Radiat Isot. 2018 Jul;137:62-67. doi: 10.1016/j.apradiso.2018.03.011. Epub 2018 Mar 14.

Abstract

The neutron autoradiography technique using polycarbonate nuclear track detectors (NTD) has been extended to quantify the boron concentration in hard tissues, an application of special interest in Boron Neutron Capture Therapy (BNCT). Chemical and mechanical processing methods to prepare thin tissue sections as required by this technique have been explored. Four different decalcification methods governed by slow and fast kinetics were tested in boron-loaded bones. Due to the significant loss of the boron content, this technique was discarded. On the contrary, mechanical manipulation to obtain bone powder and tissue sections of tens of microns thick proved reproducible and suitable, ensuring a proper conservation of the boron content in the samples. A calibration curve that relates the B concentration of a bone sample and the track density in a Lexan NTD is presented. Bone powder embedded in boric acid solution with known boron concentrations between 0 and 100 ppm was used as a standard material. The samples, contained in slim Lexan cases, were exposed to a neutron fluence of 10 cm at the thermal column central facility of the RA-3 reactor (Argentina). The revealed tracks in the NTD were counted with an image processing software. The effect of track overlapping was studied and corresponding corrections were implemented in the presented calibration curve. Stochastic simulations of the track densities produced by the products of the B thermal neutron capture reaction for different boron concentrations in bone were performed and compared with the experimental results. The remarkable agreement between the two curves suggested the suitability of the obtained experimental calibration curve. This neutron autoradiography technique was finally applied to determine the boron concentration in pulverized and compact bone samples coming from a sheep experimental model. The obtained results for both type of samples agreed with boron measurements carried out by ICP-OES within experimental uncertainties. The fact that the histological structure of bone sections remains preserved allows for future boron microdistribution analysis.

摘要

使用聚碳酸酯核径迹探测器(NTD)的中子放射自显影技术已得到扩展,用于定量硬组织中的硼浓度,这在硼中子俘获疗法(BNCT)中是一个特别感兴趣的应用。已经探索了按照该技术要求制备薄组织切片的化学和机械加工方法。在含硼骨骼中测试了四种受慢速和快速动力学控制的不同脱钙方法。由于硼含量的显著损失,该技术被舍弃。相反,通过机械操作获得骨粉和几十微米厚的组织切片被证明是可重复且合适的,确保了样品中硼含量的适当保存。给出了一条将骨样品的硼浓度与Lexan NTD中的径迹密度相关联的校准曲线。将嵌入硼浓度在0至100 ppm之间的硼酸溶液中的骨粉用作标准材料。装在细长Lexan盒中的样品在RA - 3反应堆(阿根廷)的热柱中心设施处接受10 cm的中子注量照射。用图像处理软件对NTD中显示的径迹进行计数。研究了径迹重叠的影响,并在给出的校准曲线中进行了相应校正。对不同硼浓度的骨中硼热中子俘获反应产物产生的径迹密度进行了随机模拟,并与实验结果进行了比较。两条曲线之间的显著一致性表明所获得的实验校准曲线是合适的。这种中子放射自显影技术最终被应用于测定来自绵羊实验模型的粉碎骨和致密骨样品中的硼浓度。两种类型样品获得的结果在实验不确定度范围内与通过电感耦合等离子体发射光谱法(ICP - OES)进行的硼测量结果一致。骨切片的组织结构得以保留这一事实为未来的硼微分布分析提供了可能。