Skin reactive hyperemia recorded by a combined TcPO2 and laser Doppler sensor.

作者信息

Ewald U, Huch A, Huch R, Rooth G

机构信息

Department of Paediatrics, University Hospital, Uppsala, Sweden.

出版信息

Adv Exp Med Biol. 1987;220:231-4. doi: 10.1007/978-1-4613-1927-6_42.

Abstract

The tcPO2 electrode used at 37 degrees C monitors changes in cutaneous PO2 which at this temperature is mainly determined by the changes in blood flow. Laser Doppler velocimetry (LDV) measures the product of red cell number times their velocity. A comparison of these methods for the detection of skin reactive hyperemia has been made with a combined tcPO2 and laser Doppler sensor. The two parameters increased and decreased in parallel, the tcPO2 signal being delayed 10-20 s compared to the LDV signal. A significant correlation between the two signals was obtained (r = 0.73-0.93). In contrast, no significant correlation existed between the postocclusive peaks of the two signals recorded in 12 duplicate experiments. Comparing the peak amplitudes from the repeated recordings the tcPO2 signal showed a higher correlation coefficient (r = 0.91) and smaller intercept than the LDV signal (r = 0.74). Both methods can be used for non-invasive recordings of skin reactive hyperemia. The tcPO2 signal at 37 degrees C reflects changes in blood flow of the most superficial capillaries and reactive hyperemia can be monitored with high reproducibility. Only this signal can be used for calculation of skin blood flow. The LDV signal conveys information of blood flow changes mainly of deeper dermal vascular beds and reacts virtually instantaneously, but is less reproducible.

摘要

在37摄氏度下使用的经皮氧分压(tcPO2)电极监测皮肤氧分压的变化,在此温度下,皮肤氧分压主要由血流变化决定。激光多普勒血流仪(LDV)测量红细胞数量与其速度的乘积。使用tcPO2和激光多普勒组合传感器对这些检测皮肤反应性充血的方法进行了比较。这两个参数平行增加和减少,与LDV信号相比,tcPO2信号延迟10 - 20秒。两个信号之间获得了显著相关性(r = 0.73 - 0.93)。相比之下,在12次重复实验中记录的两个信号的闭塞后峰值之间不存在显著相关性。比较重复记录的峰值幅度,tcPO2信号显示出比LDV信号更高的相关系数(r = 0.91)和更小的截距(r = 0.74)。两种方法均可用于皮肤反应性充血的无创记录。37摄氏度下的tcPO2信号反映最浅表毛细血管的血流变化,并且可以高度可重复地监测反应性充血。只有这个信号可用于计算皮肤血流量。LDV信号主要传达更深层真皮血管床血流变化的信息,并且几乎即时反应,但可重复性较差。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验