Retained percutaneous transluminal coronary angioplasty equipment components and their management.

作者信息

Hartzler G O, Rutherford B D, McConahay D R

机构信息

Mid America Heart Institute, Kansas City, Missouri.

出版信息

Am J Cardiol. 1987 Dec 1;60(16):1260-4. doi: 10.1016/0002-9149(87)90604-7.

Abstract

Of 5,400 consecutive percutaneous transluminal coronary angioplasty (PTCA) procedures, 12 patients had complications resulting in retention of 1 or more PTCA equipment components. Eight patients had guidewire fragments retained within the coronary circulation, including one with a second wire segment within the abdominal aorta. A gold band catheter marker was retained within a coronary artery in 1 patient. Four of 5 extraction procedures in these patients were successful, including retrieval of a wire segment totally contained within the distal circumflex coronary artery. Bioptomes were used to retrieve guidewire segments from the abdominal aorta in 4 patients and a knotted guiding catheter from another. At late follow-up, 5 patients with wire segment retained for an extended time within the coronary circulation had no sequelae attributable to the PTCA component debris. We conclude that many fractured intracoronary wires with proximal portion extending into the ascending aorta can be extracted. Guidewire segments retained for a long time totally within the coronary circulation may be benign, particularly when entrapped within total coronary occlusions. Bioptomes can be used effectively to remove wire segments within the abdominal aorta and to assist in the removal of kinked guide catheters.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验