The study is aimed at a verification of dose changes for a computed tomography automatic tube-current modulation (ATCM) technique. For this purpose, anthropomorphic phantom and Gafchromic XR-QA2 films were used. Radiochromic films were cut according to the shape of two thorax regions. The ATCM algorithm is based on noise index (NI) and three exam protocols with different NI were chosen, of which one was a reference. Results were compared with dose values displayed by the console and with Poisson statistics. The information obtained with radiochromic films has been normalized with respect to the NI reference value to compare dose percentage variations. Results showed that, on average, the information reported by the CT console and calculated values coincide with measurements. The study allowed verification of the dose information reported by the CT console for an ATCM technique. Although this evaluation represents an estimate, the method can be a starting point for further studies.
该研究旨在验证计算机断层扫描自动管电流调制(ATCM)技术的剂量变化。为此,使用了人体模型和Gafchromic XR-QA2薄膜。根据两个胸部区域的形状切割放射变色薄膜。ATCM算法基于噪声指数(NI),并选择了三种具有不同NI的检查方案,其中一种作为参考。将结果与控制台显示的剂量值以及泊松统计进行比较。用放射变色薄膜获得的信息已相对于NI参考值进行归一化,以比较剂量百分比变化。结果表明,平均而言,CT控制台报告的信息与计算值与测量结果一致。该研究验证了CT控制台报告的ATCM技术的剂量信息。尽管此评估只是一个估计,但该方法可作为进一步研究的起点。