Suppr超能文献

长期单中心 ICD 导线生存比较:Linox 导线早期失效的证据。

Long-term single-center comparison of ICD lead survival: Evidence for premature Linox lead failure.

机构信息

Cardiology Department, Wellington Hospital, Wellington, New Zealand.

University of Otago, Wellington, New Zealand.

出版信息

J Cardiovasc Electrophysiol. 2018 Jul;29(7):1024-1031. doi: 10.1111/jce.13502. Epub 2018 Apr 27.

Abstract

INTRODUCTION

ICD lead failure is a potential source of significant morbidity and mortality. This study investigates the survival rates of Sprint Quattro, Endotak Reliance, and Linox ICD leads.

METHODS AND RESULTS

A retrospective cohort study of all patients with an ICD implanted between January 2007 and December 2012 from the Wellington Hospital region, New Zealand, a tertiary referral center. Lead and patient details were established by review of clinical notes and the PaceArt Optima database. We analyzed a total of 287 implants using Sprint Quattro (n  =  92), Endotak Reliance (n  =  37), Linox (n  =  151), Riata (n  =  4), and Sprint Fidelis (n  =  2) leads. Median follow-up was 61.6 (26.0-81.6) months for Sprint Quattro leads, 66.7 (53.3-88.8) months for Linox leads, and 82.9 (45.9-107.4) months for Endotak Reliance leads. There were 20 cases of lead failure of which 19 were in Linox leads. The 4-, 6-, and 8-year survival for Linox leads was 97% (92.6-99.1), 93% (85.5-96.5), and 76% (62.3-85.5), respectively. The predominant abnormality was detection of nonphysiological electrical signals. Linox lead failure was associated with a younger age of patient (49.2 vs. 57.7 years, P  =  0.007).

CONCLUSIONS

The 7-year survival of Linox leads was significantly worse than shown in Biotronik surveillance reports, but in line with other single-center studies from around the world.

摘要

简介

ICD 导线故障是导致发病率和死亡率显著升高的潜在原因。本研究旨在调查 Sprint Quattro、Endotak Reliance 和 Linox ICD 导线的生存率。

方法和结果

这是一项回顾性队列研究,纳入了 2007 年 1 月至 2012 年 12 月期间在新西兰惠灵顿医院接受 ICD 植入的所有患者,该医院是一家三级转诊中心。通过审查临床记录和 PaceArt Optima 数据库确定导线和患者的详细信息。我们共分析了 287 例植入患者,其中使用 Sprint Quattro(n=92)、Endotak Reliance(n=37)、Linox(n=151)、Riata(n=4)和 Sprint Fidelis(n=2)导线。Sprint Quattro 导线的中位随访时间为 61.6(26.0-81.6)个月,Linox 导线为 66.7(53.3-88.8)个月,Endotak Reliance 导线为 82.9(45.9-107.4)个月。导线故障共有 20 例,其中 19 例发生在 Linox 导线上。Linox 导线的 4、6 和 8 年生存率分别为 97%(92.6-99.1)、93%(85.5-96.5)和 76%(62.3-85.5)。主要异常表现为检测到非生理性电信号。Linox 导线故障与患者年龄较轻相关(49.2 岁 vs. 57.7 岁,P=0.007)。

结论

Linox 导线 7 年的生存率明显差于百多力的监测报告,但与世界各地的其他单中心研究结果一致。