Department of Medical Ultrasound, Shanghai Tenth People's Hospital, Ultrasound Research and Education Institute, Tongji University School of Medicine, Shanghai, China.
Clin Hemorheol Microcirc. 2018;69(3):425-436. doi: 10.3233/CH-170334.
To evaluate a special kind of ultrasound (US) shear wave elastography for differential diagnosis of breast lesions, using a new qualitative analysis (i.e. the elasticity score in the travel time map) compared with conventional quantitative analysis.
From June 2014 to July 2015, 266 pathologically proven breast lesions were enrolled in this study. The maximum, mean, median, minimum, and standard deviation of shear wave speed (SWS) values (m/s) were assessed. The elasticity score, a new qualitative feature, was evaluated in the travel time map. The area under the receiver operating characteristic (AUROC) curves were plotted to evaluate the diagnostic performance of both qualitative and quantitative analyses for differentiation of breast lesions.
Among all quantitative parameters, SWS-max showed the highest AUROC (0.805; 95% CI: 0.752, 0.851) compared with SWS-mean (0.786; 95% CI:0.732, 0.834; P = 0.094), SWS-median (0.775; 95% CI:0.720, 0.824; P = 0.046), SWS-min (0.675; 95% CI:0.615, 0.731; P = 0.000), and SWS-SD (0.768; 95% CI:0.712, 0.817; P = 0.074). The AUROC of qualitative analysis in this study obtained the best diagnostic performance (0.871; 95% CI: 0.825, 0.909, compared with the best parameter of SWS-max in quantitative analysis, P = 0.011).
The new qualitative analysis of shear wave travel time showed the superior diagnostic performance in the differentiation of breast lesions in comparison with conventional quantitative analysis.
通过比较新的定性分析(即旅行时图中的弹性评分)与常规定量分析,评估一种特殊类型的超声(US)剪切波弹性成像技术用于鉴别诊断乳腺病变的能力。
本研究纳入了 2014 年 6 月至 2015 年 7 月期间经病理证实的 266 例乳腺病变患者。评估了剪切波速度(SWS)值(m/s)的最大值、平均值、中位数、最小值和标准差。在旅行时图中评估了一种新的定性特征,即弹性评分。绘制受试者工作特征(ROC)曲线下面积(AUROC)以评估定性和定量分析在鉴别乳腺病变方面的诊断性能。
在所有定量参数中,SWS-max 的 AUROC(0.805;95%CI:0.752,0.851)最高,与 SWS-mean(0.786;95%CI:0.732,0.834;P=0.094)、SWS-median(0.775;95%CI:0.720,0.824;P=0.046)、SWS-min(0.675;95%CI:0.615,0.731;P=0.000)和 SWS-SD(0.768;95%CI:0.712,0.817;P=0.074)相比。本研究中定性分析的 AUROC 获得了最佳的诊断性能(0.871;95%CI:0.825,0.909),与定量分析中 SWS-max 的最佳参数相比,差异具有统计学意义(P=0.011)。
与常规定量分析相比,剪切波旅行时的新定性分析在鉴别乳腺病变方面具有更好的诊断性能。