Pulsed multigated Doppler ultrasonography in the diagnosis of carotid artery disease.

作者信息

Sillesen H, Bitsch K R, Schroeder T, Steenberg H J, Hansen L, Hansen H J

机构信息

Department of Vascular Surgery, Rigshospitalet, University of Copenhagen, Denmark.

出版信息

Stroke. 1988 Jul;19(7):846-51. doi: 10.1161/01.str.19.7.846.

Abstract

To evaluate the accuracy of a pulsed multigated Doppler system, 128 carotid arteries were examined. The spectral broadening index was calculated from the power spectrum of a small sample volume located in the center of the stream according to the flow profile and was related to the degree of stenosis as determined by contrast angiography. Even minor wall irregularities seen on the angiogram were classified as disease. The ability of the system to discriminate between normal and diseased vessels reached a sensitivity of 94% and a specificity of 91%. Classification of greater than 50% or less than 50% stenosis could be performed with a sensitivity of 90% and a specificity of 85%. Pulsed multigated Doppler ultrasonography allows identification of even minor degrees of stenosis of the carotid artery and provides an alternative to duplex scanning. Furthermore, the blood flow profile provided by a multigated Doppler system may add valuable information concerning blood flow characteristics not obtainable by single-gated systems.

摘要

为评估脉冲多门控多普勒系统的准确性,对128条颈动脉进行了检查。根据血流剖面图,从小样本容积位于血流中心处的功率谱计算频谱展宽指数,并将其与经造影血管造影确定的狭窄程度相关联。血管造影上所见的即使是轻微的管壁不规则也被归类为疾病。该系统区分正常血管和病变血管的能力达到了94%的敏感性和91%的特异性。对于大于50%或小于50%的狭窄分类,敏感性为90%,特异性为85%。脉冲多门控多普勒超声检查甚至可以识别颈动脉的轻微狭窄程度,并提供了一种替代双功扫描的方法。此外,多门控多普勒系统提供的血流剖面图可能会增加有关单门控系统无法获得的血流特征的有价值信息。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验