技术说明:评估无线脉搏血氧仪在测量荷斯坦弗里生小牛动脉血氧饱和度和脉搏率中的应用。

Technical note: Evaluation of a wireless pulse oximeter for measuring arterial oxygen saturation and pulse rate in newborn Holstein Friesian calves.

机构信息

Clinical Unit for Herd Health Management in Ruminants, University Clinic for Ruminants, Department for Farm Animals and Veterinary Public Health, University of Veterinary Medicine Vienna, 1210 Vienna, Austria.

Clinical Unit for Herd Health Management in Ruminants, University Clinic for Ruminants, Department for Farm Animals and Veterinary Public Health, University of Veterinary Medicine Vienna, 1210 Vienna, Austria.

出版信息

J Dairy Sci. 2018 Jul;101(7):6437-6442. doi: 10.3168/jds.2017-14266. Epub 2018 Apr 26.

Abstract

Pulse oximetry is a well-established technique in human and veterinary medicine. In farm animals, it could also be a useful tool for the detection of critical conditions relating to oxygen supply and the cardiovascular system. Among other uses, an innovative application could be the monitoring of fetuses during birth. This could help in the early identification of critical situations and support farmers and veterinarians in their decision to start obstetric or life-support interventions. Until now, however, its use in ruminant medicine was still limited to experimental applications. The objective of this study was to evaluate the accuracy of the Radius-7 Wearable Pulse CO-Oximeter (Masimo Corporation, Irvine, CA) for monitoring vital parameters in newborn calves. All measurements were conducted on animals in the lying down position. The sensor of the pulse oximeter was placed in the interdigital space of the calves' front legs and fixed with a homemade latex hoof cover. The pulsoximetric measurements of arterial oxygen saturation (SpO) in 40 newborn calves were compared with the corresponding results (SaO) from a portable blood gas analyzer (VetScan iStat1, Abaxis Inc., Union City, CA), which served as the reference. For this, an arterial blood sample was taken from the medial intermediate branch of the caudal auricular artery. In addition, the pulse rate was measured in 10 calves aged between 0 and 7 d with the pulse oximeter and simultaneously with a heart rate belt (Polar Equine Belt, Polar Electro Oy, Kempele, Finland) to determine their level of agreement. Spearman correlation coefficient for oxygen saturation was 93.8% for the pulse oximeter and the blood gas analyzer, and 97.7% for the pulse rate measured with the pulse oximeter and the heart rate belt. Bland-Altman plots revealed an overestimation of SaO by 2.95 ± 6.39% and an underestimation of the pulse rate by -0.41 ± 3.18 beats per minute compared with the corresponding reference methods. In summary, the pulse oximeter is suitable for continuous monitoring of arterial oxygen saturation and pulse in newborn Holstein Friesian calves. For practical use, purpose-built technical equipment is required to attach the sensor to the calves' legs.

摘要

脉搏血氧仪在人类和兽医医学中是一种成熟的技术。在农场动物中,它也可能是一种有用的工具,用于检测与氧气供应和心血管系统有关的危急情况。除其他用途外,一种创新的应用可能是在分娩期间监测胎儿。这有助于早期识别危急情况,并为农民和兽医提供支持,以启动产科或生命支持干预措施。然而,到目前为止,它在反刍动物医学中的应用仍然局限于实验应用。本研究的目的是评估 Masimo 公司生产的 Radius-7 可穿戴脉搏 CO-血氧仪(Masimo Corporation,Irvine,CA)在监测新生牛犊重要参数方面的准确性。所有测量均在动物仰卧位进行。脉搏血氧仪的传感器放置在小牛前腿的指(趾)间空间,并使用自制的乳胶蹄套固定。将 40 只新生小牛的脉搏血氧饱和度(SpO)的脉搏血氧仪测量值与便携式血气分析仪(VetScan iStat1,Abaxis Inc.,Union City,CA)的相应结果(SaO)进行比较,后者作为参考。为此,从耳状中动脉的内侧中间分支采集动脉血样。此外,还使用脉搏血氧仪和心率带(Polar 马用皮带,Polar Electro Oy,Kempele,芬兰)在 10 只 0 至 7 日龄的小牛中测量心率,以确定它们的一致性水平。脉搏血氧仪和血气分析仪之间的氧饱和度的斯皮尔曼相关系数为 93.8%,脉搏血氧仪和心率带之间的心率为 97.7%。Bland-Altman 图显示,SaO 被高估了 2.95±6.39%,而与相应的参考方法相比,心率被低估了-0.41±3.18 次/分钟。综上所述,脉搏血氧仪适用于连续监测荷斯坦弗里生新生小牛的动脉血氧饱和度和脉搏。为了实际使用,需要专门的技术设备将传感器连接到小牛的腿上。