Kameda Toru, Uebayashi Kumiko, Wagai Kazuko, Kawai Fukiko, Taniguchi Nobuyuki
Department of Clinical Laboratory Medicine, Saiseikai Utsunomiya Hospital, 911-1 Takebayashi, Utsunomiya, Tochigi, 321-0974, Japan.
Department of Clinical Laboratory, Saiseikai Utsunomiya Hospital, Utsunomiya, Japan.
J Med Ultrason (2001). 2018 Oct;45(4):577-581. doi: 10.1007/s10396-018-0881-2. Epub 2018 May 2.
To assess the performance of a pocket-sized ultrasound device (PUD) for evaluating dilatation of the renal collecting system with high-end ultrasound devices (HUDs) as a reference standard.
One sonographer examined both kidneys using a PUD to evaluate dilatation of the collecting system. The grading of the dilatation ranged from 0 to 4. Immediately after the examination, another sonographer blinded to the previous results performed a formal examination with a HUD.
Two hundred kidneys in 100 patients were included in the analysis. The agreement of grades between the PUD and HUDs was excellent (weighted kappa = 0.83; P < 0.001). When hydronephrosis was defined as grade 1 or higher, the test characteristics of the PUD were as follows: sensitivity 91% (95% confidence interval (CI) 79-97%), positive predictive value 73% (95% CI 60-83%), and negative predictive value 96% (95% CI 92-99%). When hydronephrosis was defined as grade 2 or higher, the test characteristics were as follows: sensitivity 88% (95% CI 73-97%), positive predictive value 75% (95% CI 59-87%), and negative predictive value 98% (95% CI 94-99%).
Ultrasound using a PUD is useful for evaluating dilatation of the collecting system, especially for ruling out its presence.
以高端超声设备(HUD)作为参考标准,评估便携式超声设备(PUD)评估肾集合系统扩张的性能。
一名超声检查医师使用PUD检查双侧肾脏以评估集合系统扩张情况。扩张程度分级为0至4级。检查结束后,另一名对先前结果不知情的超声检查医师使用HUD进行正式检查。
分析纳入了100例患者的200个肾脏。PUD与HUD之间的分级一致性极佳(加权kappa = 0.83;P < 0.001)。当将肾积水定义为1级或更高等级时,PUD的检测特征如下:敏感性91%(95%置信区间(CI)79 - 97%),阳性预测值73%(95% CI 60 - 83%),阴性预测值96%(95% CI 92 - 99%)。当将肾积水定义为2级或更高等级时,检测特征如下:敏感性88%(95% CI 73 - 97%),阳性预测值75%(95% CI 59 - 87%),阴性预测值98%(95% CI 94 - 99%)。
使用PUD进行超声检查有助于评估集合系统扩张,尤其是用于排除其存在。