Fracture loads and failure modes of customized and non-customized zirconia abutments.

机构信息

Department of Dental Materials and Prosthodontics, Dental School of Ribeirão Preto, University of São Paulo, Avenida do Café, s/n, Monte Alegre, Ribeirão Preto, SP CEP 14040-904, Brazil.

Institute of Oral Medicine, College of Medicine, National Cheng Kung University Tainan, No. 1, Daxue Road, East District, Tainan City, Taiwan 701, Taiwan.

出版信息

Dent Mater. 2018 Aug;34(8):e197-e204. doi: 10.1016/j.dental.2018.04.005. Epub 2018 May 5.

Abstract

OBJECTIVE

This study aimed to evaluate the fracture load and pattern of customized and non-customized zirconia abutments with Morse-taper connection.

METHODS

18 implants were divided into 3 groups according to the abutments used: Zr - with non-customized zirconia abutments; Zrc - with customized zirconia abutments; and Ti - with titanium abutments. To test their load capacity, a universal test machine with a 500-kgf load cell and a 0.5-mm/min speed were used. After, one implant-abutment assembly from each group was analyzed by Scanning Electron Microscopy (SEM). For fractographic analysis, the specimens were transversely sectioned above the threads of the abutment screw in order to examine their fracture surfaces using SEM.

RESULTS

A significant difference was noted between the groups (Zr=573.7±11.66N, Zrc=768.0±8.72N and Ti=659.1±7.70N). Also, the zirconia abutments fractured while the titanium abutments deformed plastically. Zrc presented fracture loads significantly higher than Zr (p=0.009). All the zirconia abutments fractured below the implant platform, starting from the area of contact between the abutment and implant and propagating to the internal surface of the abutment. All the zirconia abutments presented complete cleavage in the mechanical test. Fractography detected differences in the position and pattern of fracture between the two groups with zirconia abutments, probably because of the different diameters in the transmucosal region.

SIGNIFICANCE

Customization of zirconia abutments did not affect their fracture loads, which were comparable to that of titanium and much higher than the maximum physiological limit for the anterior region of the maxilla.

摘要

目的

本研究旨在评估带有莫氏锥度连接的定制和非定制氧化锆基台的骨折负荷和模式。

方法

根据使用的基台,将 18 个种植体分为 3 组:Zr-使用非定制氧化锆基台;Zrc-使用定制氧化锆基台;Ti-使用钛基台。为了测试它们的承载能力,使用带有 500-kgf 负载单元和 0.5-mm/min 速度的万能试验机。之后,从每组中分析一个种植体-基台组件。为了进行断口分析,将标本在基台螺钉螺纹上方进行横向切割,以便使用扫描电子显微镜(SEM)检查其断口表面。

结果

组间差异显著(Zr=573.7±11.66N,Zrc=768.0±8.72N 和 Ti=659.1±7.70N)。此外,氧化锆基台断裂,而钛基台发生塑性变形。Zrc 的断裂载荷明显高于 Zr(p=0.009)。所有氧化锆基台均在种植体平台以下断裂,从基台和种植体之间的接触区域开始,并向基台的内表面扩展。所有氧化锆基台在机械测试中均呈现完全解理。断口分析检测到两组氧化锆基台的断裂位置和模式存在差异,这可能是由于穿粘膜区域的直径不同所致。

意义

定制氧化锆基台不会影响其断裂载荷,其与钛的断裂载荷相当,远高于上颌前区的最大生理极限。