The reliability of radiostereometric analysis in determining physeal motion in slipped capital femoral epiphysis in standard uniplanar and low-dose EOS biplanar radiography: a phantom model study.

作者信息

Hurry Jennifer K, Rehan Saad, Spurway Alan J, Laende Elise K, Astephen Wilson Janie L, Logan Karl J, Dunbar Michael J, El-Hawary Ron

机构信息

Department of Paediatric Orthopaedics, IWK Health Centre.

Department of Biomedical Engineering, Dalhousie University, Halifax, Nova Scotia, Canada.

出版信息

J Pediatr Orthop B. 2018 Nov;27(6):496-502. doi: 10.1097/BPB.0000000000000516.

Abstract

Physeal closure after slipped capital femoral epiphysis fixation can be difficult to assess on two-dimensional conventional radiographs. Radiostereometric analysis offers improved motion detection over conventional radiography, whereas the EOS biplanar imager provides a means for low radiation weight-bearing images. This phantom study assessed the reliability of measuring motion using radiostereometric analysis in the EOS using a slipped capital femoral epiphysis model. The accuracy and precision were better than 0.09±0.05 mm and 0.20°±0.36° when centered in the imaging space, were within the limits of clinical significance, and were not different from a standard uniplanar radiostereometric system.

摘要

在二维传统X线片上,股骨颈骺滑脱固定术后的骨骺闭合情况可能难以评估。与传统X线摄影相比,放射立体测量分析能更好地检测运动情况,而EOS双平面成像仪则提供了一种获取低辐射负重图像的方法。本模型研究使用股骨颈骺滑脱模型评估了在EOS中使用放射立体测量分析测量运动的可靠性。当位于成像空间中心时,其准确性和精密度优于0.09±0.05毫米和0.20°±0.36°,在临床意义范围内,且与标准单平面放射立体测量系统无差异。