Carleton University, Ottawa, Ontario, Canada.
Carleton University, Ottawa, Ontario, Canada; The Ottawa Hospital Cancer Center, Department of Medical Physics, Ottawa, Ontario, Canada.
Phys Med. 2018 Jul;51:81-90. doi: 10.1016/j.ejmp.2018.05.012.
To verify the accuracy of 4D Monte Carlo (MC) simulations, using the 4DdefDOSXYZnrc user code, in a deforming anatomy. We developed a tissue-equivalent and reproducible deformable lung phantom and evaluated 4D simulations of delivered dose to the phantom by comparing calculations against measurements.
A novel deformable phantom consisting of flexible foam, emulating lung tissue, inside a Lucite external body was constructed. A removable plug, containing an elastic tumor that can hold film and other dosimeters, was inserted in the phantom. Point dose and position measurements were performed inside and outside the tumor using RADPOS 4D dosimetry system. The phantom was irradiated on an Elekta Infinity linac in both stationary and moving states. The dose delivery was simulated using delivery log files and the phantom motion recorded with RADPOS.
Reproducibility of the phantom motion was determined to be within 1 mm. The phantom motion presented realistic features like hysteresis. MC calculations and measurements agreed within 2% at the center of tumor. Outside the tumor agreements were better than 5% which were within the positional/dose reading uncertainties at the measurement points. More than 94% of dose points from MC simulations agreed within 2%/2 mm compared to film measurements.
The deformable lung phantom presented realistic and reproducible motion characteristics and its use for verification of 4D dose calculations was demonstrated. Our 4DMC method is capable of accurate calculations of the realistic dose delivered to a moving and deforming anatomy during static and dynamic beam delivery techniques.
使用 4DdefDOSXYZnrc 用户代码,验证 4D 蒙特卡罗(MC)模拟在变形解剖中的准确性。我们开发了一种组织等效且可重复的可变形肺体模,并通过比较计算值与测量值来评估体模中传递剂量的 4D 模拟。
构建了一种由柔软泡沫组成的新型可变形体模,模仿肺组织,置于有机玻璃外部体中。体模中插入了一个可移动的插件,其中包含一个可以容纳胶片和其他剂量计的弹性肿瘤。使用 RADPOS 4D 剂量测定系统在肿瘤内部和外部进行点剂量和位置测量。体模在 Elekta Infinity 直线加速器上以固定和移动状态进行照射。使用输送日志文件模拟剂量输送,并使用 RADPOS 记录体模运动。
体模运动的可重复性被确定在 1mm 以内。体模运动呈现出真实的特征,如滞后。MC 计算值与肿瘤中心的测量值相差在 2%以内。肿瘤外部的测量值与测量点的位置/剂量读数不确定性相比,精度更好,误差在 5%以内。与胶片测量值相比,超过 94%的剂量点 MC 模拟值相差在 2%/2mm 以内。
所提出的可变形肺体模具有真实且可重复的运动特征,可用于验证 4D 剂量计算。我们的 4DMC 方法能够准确计算静态和动态射束输送技术中传递给运动和变形解剖结构的真实剂量。