Ahn H, Ivarsson L E, Johansson K, Lindhagen J, Lundgren O
Dept. of Surgery, Central Hospital, Norrköping, Sweden.
Scand J Gastroenterol. 1988 Dec;23(10):1203-10. doi: 10.3109/00365528809090192.
The use of laser Doppler flowmetry (LDF) to measure gastric blood flow was evaluated in the cat and man. The reproducibility of laser Doppler flowmetry recordings was studied in the feline stomach. In five cats flowmeter signals and venous outflow of the stomach were simultaneously recorded. The flowmeter recordings were made during operation and gastroscopy in 140 patients by means of two different (PF1-4-kHz and PF2-12-kHz) laser Doppler systems. The flowmeter recordings were highly reproducible during the cat experiments, with a coefficient of variation varying between 4% and 13%. Angulation of the probe within 60-120 degrees against the tissue under study did not affect the flowmeter signal. Pressure of the probe against the studied tissue attenuated the flowmeter signal 42 +/- 13% (n = 10). A significant correlation coefficient (r = 0.76; p = 0.01; n = 30) was obtained between flowmeter signal and venous outflow of the stomach. In man the PF2-12-kHz system yielded a higher flowmeter signal (14.5 +/- 6.9 V; 32 patients) than the PF1-4-kHz system (7.9 +/- 2.5 V; 108 patients). Flow dimensions were calculated by using results previously obtained in the intestines. The estimated blood flow values amounted to 31 +/- 10 and 57 +/- 27 ml min-1 100 g-1 for the PF1-4-kHz and PF2-12-kHz systems, respectively. In conclusion, laser Doppler flowmeter is a promising technique for studying gastric perfusion during operation and endoscopy. Motion artifacts are the major drawback of the present laser Doppler systems.
评估了激光多普勒血流仪(LDF)在猫和人体中测量胃血流量的应用。研究了猫胃中激光多普勒血流仪记录的可重复性。在五只猫中同时记录了流量计信号和胃的静脉流出量。通过两种不同的(PF1 - 4kHz和PF2 - 12kHz)激光多普勒系统,在140例患者手术和胃镜检查期间进行了流量计记录。在猫实验中,流量计记录具有高度可重复性,变异系数在4%至13%之间。探头与研究组织成60 - 120度角不会影响流量计信号。探头对研究组织的压力使流量计信号衰减42±13%(n = 10)。流量计信号与胃静脉流出量之间获得了显著的相关系数(r = 0.76;p = 0.01;n = 30)。在人体中,PF2 - 12kHz系统产生的流量计信号(14.5±6.9 V;32例患者)高于PF1 - 4kHz系统(7.9±2.5 V;108例患者)。通过使用先前在肠道中获得的结果计算血流尺寸。PFI - 4kHz和PF2 - 12kHz系统估计的血流量值分别为31±10和57±27 ml·min⁻¹·100 g⁻¹。总之,激光多普勒血流仪是一种在手术和内窥镜检查期间研究胃灌注的有前景的技术。运动伪像是当前激光多普勒系统的主要缺点。