Johnson G, Ormerod I E, Tofts P S, Barnes D, du Boulay E P
NMR Unit, Institute of Neurology, London, England.
Acta Radiol Suppl. 1986;369:496-9.
NMR relaxation time measurements derived from a set of three multi-slice sequences (which can examine the whole brain in 30 min) have been investigated. T1 and T2 values are 5% and 20% too short, respectively. Noise in the measurements was less than 2%. Temporal instability was negligible. Spatial non-uniformity was 1% (T1) and 3% (T2) within a single slice but 4% (T1) and 14% (T2) from slice to slice. Variations in cerebral tissue measurements have also been investigated: T1 varies by 6% and T2 by 12% over a population of normal controls. There can be significant patient-patient and topographical variations.
对源自一组三个多切片序列(可在30分钟内检查整个大脑)的核磁共振弛豫时间测量结果进行了研究。T1和T2值分别短了5%和20%。测量中的噪声小于2%。时间不稳定性可忽略不计。单个切片内的空间不均匀性为1%(T1)和3%(T2),但切片间为4%(T1)和14%(T2)。还对脑组织测量的变化进行了研究:在一组正常对照人群中,T1变化6%,T2变化12%。患者之间以及不同部位可能存在显著差异。