使用韦佐尼改良的巴德舍尔髋关节扩张装置进行应力位X线片时,松弛指数测量的观察者内和观察者间变异性。

Intra- and Inter-observer Variability of Measurements of the Laxity Index on Stress Radiographs Performed with the Vezzoni-Modified Badertscher Hip Distension Device.

作者信息

Bertal Mileva, Vezzoni Aldo, Houdellier Blandine, Bogaerts Evelien, Stock Emmelie, Polis Ingeborgh, Deforce Dieter, Saunders Jimmy H, Broeckx Bart J G

机构信息

Department of Medical Imaging and Small Animal Orthopaedics, Faculty of Veterinary Medicine, Ghent University, Merelbeke, Belgium.

Clinica Veterinaria Vezzoni, Cremona, Italy.

出版信息

Vet Comp Orthop Traumatol. 2018 Jul;31(4):246-251. doi: 10.1055/s-0038-1656720. Epub 2018 Jun 2.

Abstract

OBJECTIVES

To describe and evaluate the accuracy, intra- and inter-observer variability of the laxity index (LI), used to quantify hip laxity on stress radiographs obtained with the Vezzoni-modified Badertscher distension device (VMBDD).

METHODS

Stress radiographs of 10 dogs obtained with the VMBDD were measured three times by an experienced observer. Six participants with different backgrounds (two ECVDI residents, two PhD students, two veterinary assistants) followed a short presentation and performed subsequently the measurements four times in two separate sessions. The effect of self-learning, feedback and specialization on the accuracy of the measurements was assessed.

RESULTS

While the intra- and inter-observer variability were in agreement with other studies, the results of the experienced observer indicated that the variability can be very low. Neither feedback nor self-learning improved the results. A high degree of experience in radiographic assessment was not necessary to perform the measurements correctly.

CLINICAL SIGNIFICANCE

As the LI measurements were acceptable after a short presentation, they support the use of VMBDD for a complete and correct in-house evaluation of the hip joint by trained clinicians. However, we propose that, in the context of screening, measurements should be performed by a limited number of experienced examiners, to limit the impact of the inter-observer variability.

摘要

目的

描述并评估用于量化用韦佐尼改良巴德特舍尔扩张装置(VMBDD)获得的应力位X线片上髋关节松弛度的松弛指数(LI)的准确性、观察者内和观察者间的变异性。

方法

由一名经验丰富的观察者对用VMBDD获得的10只犬的应力位X线片进行三次测量。六名背景不同的参与者(两名欧洲兽医诊断成像学会住院医师、两名博士生、两名兽医助理)在听取简短介绍后,随后在两个不同的时间段内进行四次测量。评估自我学习、反馈和专业化对测量准确性的影响。

结果

虽然观察者内和观察者间的变异性与其他研究一致,但经验丰富的观察者的结果表明变异性可能非常低。反馈和自我学习均未改善结果。进行正确测量并不需要高度的X线片评估经验。

临床意义

由于在简短介绍后LI测量结果是可接受的,它们支持训练有素的临床医生使用VMBDD对髋关节进行完整且正确的内部评估。然而,我们建议,在筛查的背景下,测量应由有限数量的经验丰富的检查人员进行,以限制观察者间变异性的影响。