Suppr超能文献

中耳植入体与短突连接对中耳手术的影响

Surgical Impact of Coupling an Active Middle Ear Implant to Short Incus Process.

机构信息

Department of Otorhinolaryngology, Head and Neck Surgery "Otto Koerner", University Medical Center Rostock, Rostock.

Department of Oto-Rhino-Laryngology, Plastic, Aesthetic and Reconstructive Head and Neck Surgery, Comprehensive Hearing Center, University of Würzburg, Würzburg.

出版信息

Otol Neurotol. 2018 Jul;39(6):688-692. doi: 10.1097/MAO.0000000000001830.

Abstract

INTRODUCTION

Since 1996 the active middle-ear implant Vibrant Soundbridge (VSB) is used to treat mild-to-severe sensorineural hearing losses. The former standard surgical approach for incus vibroplasty included a mastoidectomy and a posterior tympanotomy, crimping the Floating-Mass Transducer (FMT) to the long incus process. The introduction of the short process (SP) coupler allows the attachment of the FMT to the incus body making the procedure of a posterior tympanotomy unnecessary. The aim of this study was to evaluate the duration of surgery (surgical case length) and to compare the audiological results of SP coupling with the former standard surgical approach to the long incus process.

MATERIALS AND METHODS

Patients undergoing an incus vibroplasty between October 2009 and February 2016 were included in this cohort study. The patients received an incus vibroplasty with crimping the FMT to the long incus process (former standard application) (n = 25) and with the SP coupler (n = 17) respectively. The surgical case length as well as the functional audiological outcome was assessed 12 months postoperatively using pure-tone audiometry and speech audiometry.

RESULTS

The surgical case length was significantly shorter in the SP coupler group compared to the standard application incus vibroplasty (85 ± 29 min vs. 114 ± 50 min). Additionally, patients receiving the SP coupler had an increased speech perception if compared to the standard application (Freiburger monosyllables at 65 dB SPL: 76.1 ± 16.1% vs. 66.2 ± 23.5%).

CONCLUSION

The SP coupler leads to a shortened time of surgery and by the less invasive surgery to reduced risks for facial nerve and chorda tympani. Speech perception is significantly improved by SP coupling compared to classic long incus coupling.

摘要

简介

自 1996 年以来,活跃的中耳植入体 Vibrant Soundbridge(VSB)被用于治疗轻度至重度感音神经性听力损失。以前用于镫骨振动成形术的标准手术方法包括乳突切除术和后鼓室切开术,将浮动质量换能器(FMT)压接在长镫骨突上。短突(SP)耦合器的引入允许将 FMT 连接到镫骨体上,从而使后鼓室切开术成为不必要的程序。本研究的目的是评估手术时间(手术病例长度),并比较 SP 耦合与以前的长镫骨突标准手术方法的听力结果。

材料和方法

本队列研究纳入了 2009 年 10 月至 2016 年 2 月期间接受镫骨振动成形术的患者。患者分别接受了 FMT 压接在长镫骨突上的镫骨振动成形术(以前的标准应用)(n = 25)和 SP 耦合器(n = 17)。术后 12 个月,采用纯音测听和言语测听评估手术病例长度和功能听力结果。

结果

与标准应用镫骨振动成形术相比,SP 耦合器组的手术病例长度明显缩短(85±29 分钟与 114±50 分钟)。此外,与标准应用相比,接受 SP 耦合器的患者言语感知能力提高(65 dB SPL 时的弗莱堡单音节:76.1±16.1%与 66.2±23.5%)。

结论

SP 耦合器缩短了手术时间,并通过更具侵入性的手术降低了面神经和鼓索的风险。与经典的长镫骨耦合相比,SP 耦合显著提高了言语感知能力。