通过对健眼进行二维拉伸测量法体外测试人角膜的生物力学特性:飞秒激光辅助准分子原位角膜磨镶术与小切口基质透镜切除术的比较

Biomechanical Properties of Human Cornea Tested by Two-Dimensional Extensiometry Ex Vivo in Fellow Eyes: Femtosecond Laser-Assisted LASIK Versus SMILE.

作者信息

Spiru Bogdan, Kling Sabine, Hafezi Farhad, Sekundo Walter

出版信息

J Refract Surg. 2018 Jun 1;34(6):419-423. doi: 10.3928/1081597X-20180402-05.

Abstract

PURPOSE

To investigate the biomechanical properties of the ex vivo human cornea after flap-based versus cap-based laser refractive surgery in the same donor.

METHODS

In this experimental study, 11 pairs of human corneas unsuitable for transplantation were equally divided into two groups. Corneas from the right eye were treated with femtosecond laser-assisted LASIK (FSLASIK) and corneas from the left eye with small incision lenticule extraction (SMILE). Pachymetry was measured in each eye directly before laser refractive surgery. All corneas were subjected to a refractive correction of -10.00 diopters (D) sphere and -0.75 D cylinder at 0° with a 7-mm zone, using either a 110-μm flap (FS-LASIK) or 130-μm cap (SMILE). For two-dimensional biomechanical measurements, corneoscleral buttons underwent two testing cycles (preconditioning stress-strain curve from 0.03 to 9.0 N and stress-relaxation at 9.0 N during 120 sec) to analyze the elastic and viscoelastic material properties. The effective elastic modulus was calculated. Statistical analysis was performed with a confidence interval of 95%.

RESULTS

In stress-strain measurements, the effective elastic modulus was 1.47 times higher (P = .003) after SMILE (median = 8.22 [interquartile range = 4.76] MPa) compared to FS-LASIK (median = 5.59 [inter-quartile range = 2.77] MPa). The effect size was large (r = 0.83). No significant differences (P = .658) were observed among stress-relaxation measurements, with a mean remaining stress of 181 ± 31 kPa after SMILE and 177 ± 26 kPa after FS-LASIK after relaxation.

CONCLUSIONS

Compared to a flap-based procedure such as FS-LASIK, the SMILE technique can be considered superior in terms of biomechanical stability, when measured experimentally in ex vivo human fellow eye corneas. [J Refract Surg. 2018;34(6):419-423.].

摘要

目的

研究在同一供体中,基于瓣的与基于帽的激光屈光手术后离体人角膜的生物力学特性。

方法

在本实验研究中,11对不适用于移植的人角膜被平均分为两组。右眼的角膜接受飞秒激光辅助准分子原位角膜磨镶术(FS-LASIK)治疗,左眼的角膜接受小切口基质透镜切除术(SMILE)治疗。在激光屈光手术前直接测量每只眼睛的角膜厚度。所有角膜均使用110μm瓣(FS-LASIK)或130μm帽(SMILE),在7mm区域进行-10.00屈光度(D)球镜和-0.75D柱镜、0°的屈光矫正。对于二维生物力学测量,角膜巩膜纽扣进行两个测试周期(从0.03至9.0N的预处理应力-应变曲线以及在9.0N下120秒内的应力松弛)以分析弹性和粘弹性材料特性。计算有效弹性模量。采用95%的置信区间进行统计分析。

结果

在应力-应变测量中,与FS-LASIK(中位数 = 5.59[四分位间距 = 2.77]MPa)相比,SMILE术后有效弹性模量高1.47倍(P = 0.003)(中位数 = 8.22[四分位间距 = 4.76]MPa)。效应量较大(r = 0.83)。在应力松弛测量中未观察到显著差异(P = 0.658),SMILE术后松弛后平均残余应力为181±31kPa,FS-LASIK术后为177±26kPa。

结论

与基于瓣的手术如FS-LASIK相比,当在离体人对侧眼角膜中进行实验测量时,SMILE技术在生物力学稳定性方面可被认为更优越。[《屈光手术杂志》。2018;34(6):419 - 423。]