Tahhan Nabil, Ba Papa Salmane, Hadeed Khaled, Dulac Yves, Cuttone Fabio, Leobon Bertrand
Department of Pediatric Cardiology, University Hospital, Toulouse, France.
Department of Pediatric Cardiac Surgery, University Hospital, Toulouse, France.
Ann Pediatr Cardiol. 2018 May-Aug;11(2):191-193. doi: 10.4103/apc.APC_175_17.
We report a rare and serious complication of cardiac strangulation arising from the implantation of epicardial pacing leads in a newborn. Patient's follow-up 9-month postsurgery revealed compression under the pulmonary valve annulus by a pacemaker lead, causing progressive stenosis of the right ventricular outflow tract. The epicardial leads were replaced to relieve compression, and stenosis of the right ventricular outflow tract was rectified. Pacemaker implantation in newborns is not without challenges; epicardial leads should be carefully positioned to avoid any compression of cardiac structures.