Unità Operativa di Radiologia Diagnostica ed Interventistica, IRCCS Istituto Ortopedico Galeazzi, Milan, Italy.
Dipartimento di Scienze Biomediche per la Salute, Università degli Studi di Milano, Milan, Italy.
Endocrine. 2018 Aug;61(2):336-342. doi: 10.1007/s12020-018-1647-8. Epub 2018 Jun 15.
Trabecular Bone Score (TBS) provides an indirect score of trabecular microarchitecture from lumbar spine (LS) dual energy X-ray absorptiometry. Increasing soft tissue thickness artifactually reduces TBS values; we evaluated the effect of a fictitious increase of soft tissue thickness on TBS and bone mineral density (BMD) reproducibility on a phantom model.
A Hologic spine phantom was scanned with a QDR-Discovery W Hologic densitometer. Fresh pork rind layers of 5 mm were used to simulate the in-vivo soft tissues. For each scan mode (fast array [FA], array, high definition [HD]), 25 scans were consecutively performed without phantom repositioning, at 0 (no layers), 1 cm, 3 cm, and 6 cm of thickness. BMD and TBS reproducibility was calculated as the complement to 100% of least significant change.
Both BMD and TBS reproducibility slightly decreased with increasing soft tissue; this difference was statistically significant only for BMD using HD modality (reproducibility decreased from 99.4% at baseline to 98.4% at 6-cm of thickness). TBS reproducibility was slightly lower compared to that of BMD, and ranged between 98.8% (array, 0 cm) and 97.4% (FA, 6 cm). Without taking into account manufacturer BMI optimization, we found a progressive decrease of TBS mean values with increasing soft tissue thickness. The highest TBS difference between baseline scan and 6 cm was -0.179 (-14.27%) using HD.
Despite being slightly lower than that of BMD, TBS reproducibility was not affected up to 6 cm of increasing soft tissue thickness, and was even less influenced by fat than BMD reproducibility.
小梁骨评分(TBS)提供了一种从腰椎(LS)双能 X 射线吸收法间接评估小梁微观结构的评分。软组织厚度的增加会人为地降低 TBS 值;我们评估了在体模模型上人为增加软组织厚度对 TBS 和骨密度(BMD)可重复性的影响。
使用 Hologic 脊柱体模在 QDR-Discovery W Hologic 密度仪上进行扫描。新鲜猪皮层厚度为 5mm,用于模拟体内软组织。对于每个扫描模式(快速阵列[FA]、阵列、高清晰度[HD]),在不重新定位体模的情况下,连续进行 25 次扫描,厚度分别为 0(无层)、1cm、3cm 和 6cm。BMD 和 TBS 重复性的计算方法是将最小有意义变化的 100%的补数。
随着软组织厚度的增加,BMD 和 TBS 的重复性都略有下降;仅在使用 HD 模式时,BMD 的差异具有统计学意义(重复性从基线时的 99.4%下降至 6cm 时的 98.4%)。与 BMD 相比,TBS 的重复性略低,范围在 98.8%(阵列,0cm)至 97.4%(FA,6cm)之间。在不考虑制造商 BMI 优化的情况下,我们发现 TBS 平均值随软组织厚度的增加呈逐渐下降趋势。在使用 HD 时,基线扫描和 6cm 之间的 TBS 差值最高为-0.179(-14.27%)。
尽管略低于 BMD,但 TBS 重复性在增加 6cm 软组织厚度的情况下不受影响,并且受脂肪影响甚至比 BMD 重复性更小。