High-resolution 3D mapping of epicardial conduction during Marshall bundle-related atrial tachycardia.

作者信息

Hojo Rintaro, Fukamizu Seiji, Miyazawa Satoshi, Kawamura Iwanari

机构信息

Department of Cardiology Tokyo Metropolitan Hiroo Hospital Tokyo Japan.

出版信息

J Arrhythm. 2018 Apr 30;34(3):298-301. doi: 10.1002/joa3.12067. eCollection 2018 Jun.

Abstract

A 73-year-old woman was admitted for atrial tachycardia (AT) ablation. The activation map and pacing study indicated that the AT propagated around the left pulmonary vein and that the Marshall bundle (MB) bypassed the scar area of the left pulmonary vein ridge and mitral isthmus. The Rhythmia Mapping System revealed double potentials propagated along the assumed position of the MB. The mapping system includes a confidence mask that can be used to visually identify low-confidence areas of the map based upon extremely low-voltage signals. Given the low-voltage area in the endocardial side, the epicardial conduction was emphasized.

摘要

一名73岁女性因房性心动过速(AT)消融入院。激动标测和起搏研究表明,AT沿左肺静脉周围传导,而马歇尔束(MB)绕过左肺静脉嵴和二尖瓣峡部的瘢痕区域。Rhythmia标测系统显示双电位沿MB的假定位置传导。该标测系统包括一个置信掩码,可用于根据极低电压信号直观识别标测图中的低置信区域。鉴于心内膜侧存在低电压区域,因此强调了心外膜传导。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d4cc/6009767/ec7a66bfff34/JOA3-34-298-g001.jpg