Suppr超能文献

用于一体式陶瓷种植体植入的无袖三维钻孔导向器的精度:一项前瞻性临床试验。

Precision of sleeveless 3D drill guides for insertion of one-piece ceramic implants: a prospective clinical trial.

作者信息

Schnutenhaus Sigmar, von Koenigsmarck Valerie, Blender Sarah, Ambrosius Laura, Luthardt Ralph G, Rudolph Heike

出版信息

Int J Comput Dent. 2018;21(2):97-105.

Abstract

OBJECTIVE

The insertion of implants through template-guided surgery is an established method. At the present time, the static navigation is based on sleeves integrated into the guides, through which drill bits of the corresponding size are passed. The present clinical trial is designed to test a new sleeveless system to determine the precision of implant insertion for one-piece ceramic implants.

MATERIALS AND METHODS

For 12 patients, implant bed preparation and the insertion of the implants were done using a sleeveless implant guide. Twenty implants were inserted in this way and checked for precision. The check was done using a noninvasive method, which permitted comparison of the planning data with the actually realized positions after superimposition.

RESULTS

The mean deviations were 0.52 mm (95% CI: 0.37 to 0.67 mm) at the crestal position of the implant, and 0.82 mm (95% CI: 0.56 to 1.08 mm) at the apical tip of the implant. The height deviation was 0.35 mm (95% CI: 0.01 to 0.68 mm). The mean angular deviation was measured as 2.85 degrees (95% CI: 2.18 to 3.51 degrees).

CONCLUSION

The values showed good precision in all the parameters measured. The results were thus in a range equal to or better than the mean precision found in numerous clinical trials described in the literature. In particular, the method showed little deviation, as illustrated by the confidence interval (CI), but also by the clinically critical maximum deviations.

摘要

目的

通过模板引导手术植入种植体是一种既定的方法。目前,静态导航基于集成在导向器中的套筒,相应尺寸的钻头通过该套筒。本临床试验旨在测试一种新的无套筒系统,以确定一体式陶瓷种植体植入的精度。

材料与方法

对12名患者使用无套筒种植体导向器进行种植床准备和种植体植入。以这种方式植入了20颗种植体并检查其精度。检查采用非侵入性方法,该方法允许在叠加后将规划数据与实际实现的位置进行比较。

结果

种植体嵴顶位置的平均偏差为0.52毫米(95%置信区间:0.37至0.67毫米),种植体根尖处的平均偏差为0.82毫米(95%置信区间:0.56至1.08毫米)。高度偏差为0.35毫米(95%置信区间:0.01至0.68毫米)。平均角度偏差测量为2.85度(95%置信区间:2.18至3.51度)。

结论

这些值在所有测量参数中显示出良好的精度。因此,结果处于等于或优于文献中描述的众多临床试验中发现的平均精度的范围内。特别是,该方法显示出很小的偏差,这不仅由置信区间(CI)表明,也由临床关键的最大偏差表明。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验