Hemodynamic assessments of venous pulsatile tinnitus using 4D-flow MRI.

机构信息

From the Center for Biomedical Imaging Research (Y.L., H.C., L.H., R.L., C.Y.), Department of Biomedical Engineering, Tsinghua University; Neurosurgery Department (X.C.), PLA General Hospital; Xuanwu Hospital (X.W.), Capital Medical University; Beijing Tongren Hospital (S.C.), Capital Medical University, Beijing, China; and Department of Radiology (C.Y.), University of Washington, Seattle.

出版信息

Neurology. 2018 Aug 7;91(6):e586-e593. doi: 10.1212/WNL.0000000000005948. Epub 2018 Jul 11.

Abstract

OBJECTIVE

To use 4D-flow MRI to characterize hemodynamics of transverse and sigmoid sinus in venous pulsatile tinnitus (PT) patients and to investigate their differences vs healthy controls.

METHODS

A total of 21 patients with venous PT and 11 healthy controls were included in the retrospective study. All participants underwent 4D-flow and magnetic resonance venography scan in a 3.0T magnetic resonance scanner. All visualization, quantification, and analysis of 4D-flow data were performed using dedicated software. Two independent reviewers evaluated the existence of vortex or turbulence. Covariance analysis adjusted for age was used to compare average through-plane velocity (V), maximum through-plane velocity (V), average velocity (V), maximum velocity (V), average blood flow (Flow), and pulsatility index (PI) between PT and control group.

RESULTS

There were hemodynamic differences between PT patients and healthy controls. Compared with the control group, the PT group showed significantly higher V, V, V, V, and Flow, and slightly higher PI. For the assessment of flow pattern, inter-reader reproducibility was excellent (κ = 1.00). Vortex or turbulence was observed in PT patients with good sensitivity (86.4%) and specificity (90.9%). Drainage dominance was more frequently observed in patients (15/21, 71.4%) than healthy controls (4/11, 36.4%).

CONCLUSIONS

Significant hemodynamic differences were found between venous PT patients and healthy controls with 4D-flow MRI. Hemodynamic conditions could serve as noninvasive biomarkers for diagnosis and treatment evaluation of venous PT.

CLASSIFICATION OF EVIDENCE

This study provides Class III evidence that 4D-flow MRI accurately identifies patients with venous PT.

摘要

目的

使用 4D-flow MRI 对搏动性耳鸣(PT)患者的横窦和乙状窦血流动力学进行特征分析,并研究其与健康对照组的差异。

方法

本回顾性研究共纳入 21 例静脉性 PT 患者和 11 名健康对照者。所有参与者均在 3.0T 磁共振扫描仪上接受 4D-flow 和磁共振静脉造影扫描。使用专用软件对 4D-flow 数据进行可视化、定量和分析。两名独立的观察者评估涡流或湍流的存在。采用协方差分析调整年龄进行比较,以比较 PT 组和对照组之间的平面内平均速度(V)、平面内最大速度(V)、平均速度(V)、最大速度(V)、平均血流(Flow)和搏动指数(PI)。

结果

PT 患者与健康对照组之间存在血流动力学差异。与对照组相比,PT 组的 V、V、V、V 和 Flow 显著更高,PI 略高。对于血流模式的评估,两位观察者之间的重复性极好(κ=1.00)。PT 患者中观察到涡流或湍流,具有良好的敏感性(86.4%)和特异性(90.9%)。与健康对照组(4/11,36.4%)相比,患者中更常观察到引流优势(15/21,71.4%)。

结论

4D-flow MRI 发现静脉性 PT 患者与健康对照组之间存在显著的血流动力学差异。血流动力学条件可作为静脉性 PT 诊断和治疗评估的无创生物标志物。

证据分类

本研究提供了 III 级证据,表明 4D-flow MRI 可准确识别静脉性 PT 患者。