Raffaeli Genny, Canesi Francesco, Conigliaro Federica, Ghirardello Stefano, Vanzati Mara, Baracetti Chiara, Fumagalli Monica, Ciralli Fabrizio, Schena Federico, Pesenti Nicola, Plevani Laura, Mosca Fabio, Cavallaro Giacomo
NICU, Fondazione IRCCS Cà Granda Ospedale Maggiore Policlinico, Università degli Studi di Milano, Milano, Italy.
NICU, Fondazione IRCCS Cà Granda Ospedale Maggiore Policlinico, Università degli Studi di Milano, Milano, Italy; Betamed Perfusion Service, Rome, Italy.
Heart Lung. 2018 Nov;47(6):638-645. doi: 10.1016/j.hrtlng.2018.06.007. Epub 2018 Aug 3.
Advances in cardiorespiratory monitoring have made the extracorporeal membrane oxygenation (ECMO) technique safer for the patient. Noninvasive, continuous tools are available, although data on their applications in the neonatal ECMO setting are lacking.
We retrospectively described the neonatal clinical application of this continuous, noninvasive ECMO monitor and compared the analyzed parameters from those derived from blood gas analysis.
We performed 897 h of cardiorespiratory monitoring during neonatal venoarterial-ECMO (VA-ECMO) for four patients affected by (cardio-) respiratory failure, to compare the reliability of a noninvasive, continuous monitoring Spectrum M4® (M4) (Spectrum Medical, Gloucester, England) to an invasive, intermittent co-monitoring with blood gas analyzer (Radiometer Medical ApS, Brønshøj, Denmark).
A range of 117 pairs (time-matched BGA-derived vs. M4-derived parameters) was retrospectively analyzed. T-test, linear regression and Bland-Altman analysis for hemoglobin, hematocrit, venous oxygen saturation, oxygen delivery, oxygen consumption, oxygen extraction ratio, oxygen partial pressure, and carbon dioxide partial pressure showed a strong relationship between the two monitors for all parameters analyzed (p < 0.0001).
Continuous, noninvasive cardiorespiratory monitoring appears to be feasible and reliable, although its accuracy needs to be verified in a more extensive cohort.