: Gaitography is gait parametrization from center-of-pressure trajectories of walking on an instrumented treadmill. Gaitograms may be useful for prosthetic gait analyses, as they can be rapidly and unobtrusively collected over multiple gait cycles without constraining foot placement. However, its reliability must still be established for prosthetic gait.
: To evaluate (a) within-method test-retest repeatability and (b) between-methods agreement for temporal gait events (foot contact, foot off) and gait characteristics (e.g. step times, single-support duration).
: Cohort study with repeated measurements.
: Ten male proficient prosthetic walkers with a unilateral trans-femoral or trans-tibial amputation were equipped with a pressure-insole system and were invited to walk on separate days on an instrumented treadmill.
: We found better between-methods reproducibility than within-method repeatability in temporal gait characteristics. Step times, stride times, and foot-contact events matched well between the two methods. In contrast, insole-based foot-off events were detected one-to-two samples earlier. Likewise, a similar bias was observed for temporal gait characteristics that incorporated foot-off events.
: Notwithstanding small systematic biases, the good between-methods agreement indicates that temporal gait characteristics may be determined interchangeably with gaitograms and insoles in persons with a prosthesis. However, the relatively poorer test-retest repeatability hinders longitudinal assessments with either method.
: Clinical practice could potentially benefit from gaitography as an efficient, unobtrusive, easy to use, automatized, and patient-friendly means to objectively parametrize prosthetic gait, with immediate availability of test results allowing for prompt clinical decision-making. Temporal gait parameters demonstrate good between-methods agreement, but poorer within-method repeatability hinders detecting prosthetic gait changes.
步态描记法是根据在装有仪器的跑步机上行走时的压力中心轨迹对步态进行参数化。步态图可能对假肢步态分析有用,因为它们可以在多个步态周期中快速且不显眼地收集,而不会限制足部位置。然而,其在假肢步态方面的可靠性仍有待确定。
评估(a)方法内重测重复性以及(b)时间步态事件(足接触、足离地)和步态特征(如步时、单支撑持续时间)的方法间一致性。
重复测量的队列研究。
十名单侧经股骨或经胫骨截肢的男性熟练假肢步行者配备了压力鞋垫系统,并被邀请在不同日期在装有仪器的跑步机上行走。
我们发现,在时间步态特征方面,方法间的可重复性优于方法内重复性。两种方法之间的步时、步幅时间和足接触事件匹配良好。相比之下,基于鞋垫的足离地事件检测提前了一到两个样本。同样,在纳入足离地事件的时间步态特征方面也观察到了类似的偏差。
尽管存在小的系统偏差,但良好的方法间一致性表明,假肢使用者的时间步态特征可以用步态图和鞋垫交替确定。然而,相对较差的重测重复性阻碍了使用这两种方法进行纵向评估。
临床实践可能会从步态描记法中受益,它是一种客观地对假肢步态进行参数化的高效、不显眼、易于使用、自动化且对患者友好的方法,测试结果可立即获得,有助于迅速做出临床决策。时间步态参数显示出良好的方法间一致性,但较差的方法内重复性阻碍了对假肢步态变化的检测。