Suppr超能文献

扭转预加载后镍钛控制记忆和蓝色技术器械的循环弯曲疲劳抗性

Cyclic flexural fatigue resistance of NiTi Controlled Memory and Blue Technology instruments after torsional preloading.

作者信息

Modesto Túlio César, Acosta Eufemia Carolina Peláez, Resende Pedro Damas, Pereira Érika Sales Joviano, Peixoto Isabella Faria da Cunha, Buono Vicente Tadeu Lopes, Viana Ana Cecília Diniz

机构信息

Departmento de Odontologia Restauradora, Faculdade de Odontologia, Universidade Federal de Minas Gerais, Belo Horizonte, MG, Brasil.

Departmento de Engenharia Metalúrgica e de Materiais, Faculdade de Engenharia, Universidade Federal de Minas Gerais, Belo Horizonte, MG, Brasil.

出版信息

J Appl Oral Sci. 2018;26:e20180144. doi: 10.1590/1678-7757-2018-0144. Epub 2018 Aug 20.

Abstract

Objective The aim of this study was to evaluate the influence of torsional preloading on the cyclic flexural fatigue resistance of thermally treated NiTi instruments. Material and Methods Ten new instruments New Hyflex CM (HF 30.06; Coltene/Whaladent Inc.), Typhoon CM (TYP 30.06; Clinician's Choice Dental Products) and Vortex Blue (VB 30.06; Dentsply Tulsa Dental) were chosen, based on geometry and specific characteristics of the manufacturing process. The new instruments of each system were tested in a bench device to determine their fatigue resistance through mean value of number of cycles to failure (Nf) (Control Group - CG). Another group of 10 new HF, TYP and VB instruments were submitted to 20 cycles of torsional straining between 0° and 180° (Experimental Group - EG) and then submitted to fatigue until rupture under the same conditions of the CG. Tested instruments were examined by scanning electron microscopy (SEM). Data were analyzed using one-way analysis of variance and post hoc Tukey's test (α=.05). Results Higher fatigue resistance was accomplished by HF instruments, followed by VB and TYP (p<0.05). During the torsional preloading, the lowest mean torque value was observed for TYP instruments (p<0.05). The torsional preload caused a significant reduction in the Nf values (p<0.05) of about 20%, 39% and 45% for instruments HF, VB and TYP, respectively. Longitudinal cracks, generated during the torsional preloading, were present in VB files, but were not observed in the CM instruments (HF and TYP). Conclusions In conclusion, the flexural fatigue resistance of thermally treated instruments is diminished after cyclic torsional loading.

摘要

目的 本研究旨在评估扭转预加载对热处理镍钛器械循环弯曲疲劳抗力的影响。

材料与方法 根据制造工艺的几何形状和特定特性,选择了10支新型器械:New Hyflex CM(HF 30.06;Coltene/Whaladent公司)、Typhoon CM(TYP 30.06;Clinician's Choice Dental Products公司)和Vortex Blue(VB 30.06;Dentsply Tulsa Dental公司)。每个系统的新型器械在台架装置中进行测试,通过失效循环次数(Nf)的平均值来确定其疲劳抗力(对照组 - CG)。另一组10支新的HF、TYP和VB器械在0°至180°之间进行20次扭转应变循环(实验组 - EG),然后在与CG相同的条件下进行疲劳测试直至断裂。对测试后的器械进行扫描电子显微镜(SEM)检查。数据采用单因素方差分析和事后Tukey检验(α = 0.05)进行分析。

结果 HF器械具有更高的疲劳抗力,其次是VB和TYP(p < 0.05)。在扭转预加载过程中,TYP器械的平均扭矩值最低(p < 0.05)。扭转预加载使HF、VB和TYP器械的Nf值分别显著降低约20%、39%和45%(p < 0.05)。在扭转预加载过程中产生的纵向裂纹出现在VB锉中,但在CM器械(HF和TYP)中未观察到。

结论 总之,循环扭转加载后,热处理器械的弯曲疲劳抗力降低。