Department of Radiation Oncology, Universitätsmedizin Mannheim, University of Heidelberg, Mannheim, Germany.
Department of Radiation Oncology, Universitätsmedizin Mannheim, University of Heidelberg, Mannheim, Germany.
Phys Med. 2018 Aug;52:93-97. doi: 10.1016/j.ejmp.2018.06.640. Epub 2018 Jul 6.
Frame based positioning accuracy in Gamma Knife (GK) stereotactic radiosurgery (SRS) is extremely high but removal of a post may be necessary to enable the treatment in selected patients.
To verify the positioning accuracy in clinical scenarios with 4 and 3 posts in patients and phantoms using cone-beam CT (CBCT) of Gamma Knife Icon™.
We analyzed positioning accuracy for 12 patients with standard 4 post setup using pretreatment CBCT (pre-CBCT) on GK Icon™ and report 4 patients with different clinical scenarios (removal of a post). We performed phantom measurements to verify the frame accuracy via CBCT in different clinical scenarios without the influence of the human patient.
Mean frame accuracy for 12 patients with 4 posts was 0.35 mm/0.34 degree. Mean motion during treatment was 0.11 mm/0.04 degree. For two of the clinical scenarios where a post was removed, we found acceptable deviations within 0.66 mm/0.61 degree. For 2 patients, a deviation of 2.94 mm/-3.47 degree and 1.85 mm/-0.74 degree was found and replanning was necessary. Phantom measurements showed good agreement when planning MR/CT was performed with 4 or 3 post. Larger deviations of 0.86 mm/0.88 degree were detected when a post was removed after planning MR/CT.
The frame accuracy with 4 posts before and during GK treatments is as high as expected. For clinical situations, where a post is removed after planning-CT/MR, pre-treatment position verification is strongly suggested using stereotactic CBCT or the P-CT/MR should be repeated to avoid possible mistreatments.
伽玛刀(GK)立体定向放射外科(SRS)的框架定位精度极高,但在某些选定的患者中,可能需要移除一根支柱以进行治疗。
使用伽玛刀 Icon™的锥形束 CT(CBCT)验证 4 根和 3 根支柱的患者和体模的临床情况下的定位精度。
我们分析了 12 名使用 GK Icon™上的预处理 CBCT(pre-CBCT)进行标准 4 根支柱设置的患者的定位精度,并报告了 4 名具有不同临床情况(移除一根支柱)的患者。我们进行了体模测量,以在没有患者影响的不同临床情况下通过 CBCT 验证框架准确性。
对于 12 名具有 4 根支柱的患者,平均框架精度为 0.35mm/0.34 度。治疗过程中的平均运动为 0.11mm/0.04 度。对于两个移除一根支柱的临床情况,我们发现可接受的偏差在 0.66mm/0.61 度内。对于 2 名患者,发现偏差为 2.94mm/-3.47 度和 1.85mm/-0.74 度,需要重新计划。当使用 4 根或 3 根支柱进行规划的 MR/CT 时,体模测量显示出良好的一致性。在计划的 MR/CT 后移除一根支柱时,检测到较大的偏差为 0.86mm/0.88 度。
在 GK 治疗前后使用 4 根支柱的框架精度与预期相符。对于计划 CT/MR 后移除一根支柱的临床情况,强烈建议使用立体定向 CBCT 进行治疗前位置验证,或重复进行 P-CT/MR,以避免可能的误治疗。