McLeod Christopher, Moran Carmel M, McBride Karne A, Pye Stephen D
Medical Physics, NHS Lothian, Edinburgh Royal Infirmary, Edinburgh, United Kingdom.
Centre for Cardiovascular Science, University of Edinburgh, Edinburgh, United Kingdom.
Ultrasound Med Biol. 2018 Dec;44(12):2802-2812. doi: 10.1016/j.ultrasmedbio.2018.07.014. Epub 2018 Aug 23.
Intravascular ultrasound (IVUS) catheters are a specialist imaging modality used in the assessment of cardiovascular disease. The ultrasound transducer may either be of single-element mechanical or phased-array design. Because of their design and operating frequencies (10-45 MHz), evaluation of the imaging performance is not possible with commercially available ultrasound test objects. An existing test object, the Edinburgh Pipe Phantom, was modified to allow measurement of resolution integral (R), depth of field (Lr) and characteristic resolution (Dr) of IVUS catheters. In total, seven IVUS catheters, from two manufacturers and of both single-element mechanical and phased-array design, were tested to provide a measure of performance over different frequencies and technologies. Measurements of R for the tested IVUS catheters ranged from 11.9 to 18.8. The modified Edinburgh Pipe Phantom therefore allows catheter-based ultrasound probes to be evaluated scientifically and their performance to be seen in relation to other similar ultrasound technologies such as pre-clinical ultrasound and endoscopic ultrasound.
血管内超声(IVUS)导管是用于评估心血管疾病的一种专业成像方式。超声换能器可以是单元素机械设计或相控阵设计。由于其设计和工作频率(10 - 45兆赫兹),使用市售超声测试物体无法评估成像性能。对现有的测试物体——爱丁堡管道模型进行了修改,以允许测量IVUS导管的分辨率积分(R)、景深(Lr)和特征分辨率(Dr)。总共测试了来自两家制造商的七个IVUS导管,包括单元素机械设计和相控阵设计,以衡量不同频率和技术下的性能。测试的IVUS导管的R测量值范围为11.9至18.8。因此,经过修改的爱丁堡管道模型允许对基于导管的超声探头进行科学评估,并能将其性能与其他类似的超声技术(如临床前超声和内镜超声)进行比较。