Physiol Meas. 2018 Sep 27;39(9):098001. doi: 10.1088/1361-6579/aadf11.
A recent study investigated the potential utility of new photoplethysmogram (PPG) indicators to improve cuffless continuous blood pressure (BP) measurement.
In this Comment we provide additional discussion of the physiology underlying PPG- and pulse arrival time (PAT)-based BP measurement. We also discuss validation of these measurements.
Changes in PPG features and PAT can occur independently of BP. Any study validating BP measurement based on PPG features or PAT should include a variety of calibration independent physiological challenges affecting BP.
PPG/PAT-based BP measurement has been investigated extensively as an alternative to cuff-based BP measurement. We feel the inherent physiological confounding in PPG/PAT-based measurement makes it unlikely that it will be of clinical value.