非骨水泥、弯曲、短柄股骨远端重建内假体柄:早期随访结果。

Uncemented, curved, short endoprosthesis stem for distal femoral reconstruction: early follow-up outcomes.

机构信息

Department of Orthopedics, West China Hospital, Sichuan University, No. 37 Guoxue Street, Chengdu, 610041, People's Republic of China.

Department of Orthopedics, The Third Hospital of Mianyang, No. 190 The East Jiannan Road, Mianyang, 621000, Sichuan, People's Republic of China.

出版信息

World J Surg Oncol. 2018 Sep 10;16(1):183. doi: 10.1186/s12957-018-1486-3.

Abstract

BACKGROUND

Uncemented endoprosthetic knee replacement has become a mainstream treatment for malignant tumours of the distal femur. Most femoral stems, however, are straight and therefore poorly fit the anteriorly bowed curvature of the femur. To address this issue, we used a short, curved, uncemented press-fit femoral stem and evaluated its short-term outcomes after reconstruction of the distal femur.

METHODS

Forty-two patients underwent distal femur replacement using curved press-fit stem. To assess the interface, we measured the axial length of the press-fit area and the perpendicular distance of the radiolucent area between the stem and bone on digital images obtained using tomosynthesis with Shimadzu Metal Artefact Reduction Technology (T-SMART). Postoperative complications and oncological outcomes were monitored at each follow-up visit.

RESULTS

Of the 42 patients enrolled in the study, two had cancer-related deaths and one had local tumour recurrence. The minimum follow-up time of the surviving patients was 24 months, with no incidence of aseptic loosening or mechanical failure of the prosthesis. The average effective contact length between the press-fit stem and bone was 74.0 mm, with nearly undetectable radiolucent gaps between the implant and the bone on medial-lateral and anteroposterior views.

CONCLUSIONS

Over the short term, uncemented, curved, short stem provides a stable bone-prosthesis interface without any aseptic loosening.

摘要

背景

非骨水泥型人工膝关节假体置换术已成为治疗股骨远端恶性肿瘤的主流方法。然而,大多数股骨柄为直型,与股骨的前弓状曲率适配不佳。为了解决这个问题,我们使用了一种短的、弯曲的、非骨水泥压配合股骨柄,并评估了其在股骨远端重建后的短期结果。

方法

42 例患者采用弯曲压配合股骨柄进行股骨远端置换。为了评估界面,我们在使用 Shimadzu 金属伪影减少技术(T-SMART)的断层合成数字图像上测量压配合区域的轴向长度和柄与骨之间的透明区的垂直距离。在每次随访时监测术后并发症和肿瘤学结果。

结果

在纳入研究的 42 例患者中,有 2 例因癌症相关死亡,1 例局部肿瘤复发。存活患者的最短随访时间为 24 个月,无假体无菌性松动或机械故障的发生。压配合柄与骨之间的平均有效接触长度为 74.0mm,在内外侧和前后位视图上,植入物与骨之间几乎没有可检测到的透明间隙。

结论

在短期内,非骨水泥、弯曲、短柄提供了一个稳定的骨-假体界面,没有任何无菌性松动。