Zhang Lingyun, Jiang Yugang
Neurosurgery Department, The Second Xiangya Hospital of Central South University, Changsha, Hunan 410011, P.R. China.
Exp Ther Med. 2018 Sep;16(3):2627-2632. doi: 10.3892/etm.2018.6503. Epub 2018 Jul 23.
The present study aimed to evaluate the operation, application and safety of a novel prototype device designed for the retrieval of a migrated coils, an intraprocedural complication of endovascular aneurysm treatment, compared with the widely used Solitaire AB stent retriever. The prototype retrieval device, referred to as magnetic wires, consisted of two microwires that were joined by a small magnetic ring near to the tips. To retrieve a coil, the wires are advanced through the dislodged coil one at a time and then capture the target coil in a chopstick-like manner, as their tips are attracted to the magnet. tests were performed five times for each retrieval device using a pulsing flow circulator to compare retrieval effectiveness. efficacy testing was performed in 5 male piglets. The left and right carotid arteries were used to assess the magnetic wires and the Solitaire retriever, respectively. While the two devices successfully retrieved coils at a rate of 100%, the magnetic wires did so in a shorter time. , the magnetic wires successfully captured the migrated coils in 3/5 cases (60%), whereas the Solitaire AB stent retriever was successful 4/5 cases (80%). Endothelial injury was observed following magnetic wires retrieval in 1/5 cases (20%) and following use of the Solitaire AB stent retriever in 3/5 cases (60%). These results demonstrate the feasibility and safety of this novel prototype coil retrieval device, which with further development has the potential to be an optimal device for the retrieval of dislodged coils.
本研究旨在评估一种新型原型设备的操作、应用及安全性,该设备专为取出移位的弹簧圈而设计,移位弹簧圈是血管内动脉瘤治疗的一种术中并发症,将其与广泛使用的Solitaire AB支架取栓器进行比较。这种原型取栓设备称为磁丝,由两根微丝组成,在靠近尖端处通过一个小磁环连接。为了取出弹簧圈,将磁丝一次一根穿过移位的弹簧圈,然后以筷子夹物的方式捕获目标弹簧圈,因为它们的尖端会被磁体吸引。使用脉冲流循环器对每个取栓设备进行了5次测试,以比较取栓效果。在5只雄性仔猪身上进行了疗效测试。分别使用左右颈动脉来评估磁丝和Solitaire取栓器。虽然两种设备均以100%的成功率取出了弹簧圈,但磁丝用时更短。此外,磁丝在3/5例(60%)中成功捕获了移位的弹簧圈,而Solitaire AB支架取栓器在4/5例(80%)中成功。在1/5例(20%)磁丝取栓后和3/5例(60%)使用Solitaire AB支架取栓器后观察到了内皮损伤。这些结果证明了这种新型原型弹簧圈取栓设备的可行性和安全性,随着进一步发展,它有可能成为取出移位弹簧圈的最佳设备。