Suppr超能文献

电子近距离放射治疗皮肤癌时有效治疗深度的评估

Evaluation of effective treatment depth in skin cancer treatments with electronic brachytherapy.

作者信息

Lloyd Samantha A M, Rahn Douglas A, Hoisak Jeremy D P, Dragojević Irena

机构信息

Department of Radiation Medicine and Applied Sciences, University of California San Diego, La Jolla, CA.

Department of Radiation Medicine and Applied Sciences, University of California San Diego, La Jolla, CA.

出版信息

Brachytherapy. 2018 Nov-Dec;17(6):990-994. doi: 10.1016/j.brachy.2018.08.003. Epub 2018 Sep 11.

Abstract

PURPOSE

To evaluate changes in the percent depth dose (PDD) and effective depth of treatment due to force applied by the applicator during treatments of nonmelanoma skin cancer with the Xoft electronic brachytherapy system.

METHODS

To simulate compressible tissue, a 5-mm tissue-equivalent bolus was used. A soft x-ray ion chamber was used for output measurements, which were performed for all Xoft surface applicators with plastic endcaps in place. Output was first measured at 5 mm depth with minimal pressure from the applicator on the bolus and then repeated after applying uniform pressure on the applicator to calculate the change in PDD and effective treatment depth.

RESULTS

For the 10-mm cone, a moderate force of 5 N changed the PDD by more than 20%. The effect was also pronounced for the 20-mm cone, while minimal for the 35- and 50-mm cones. Even when only a moderate force was applied, the effective prescription depth changed by several millimeters, on the order of a typical prescription depth.

CONCLUSION

Based on the results of this simulation, excessive pressure applied on the skin by the applicator can drastically alter the PDD and effective treatment depth. The effect is most pronounced for the 10- and 20-mm cones, which tend to be used most frequently. Inappropriate applicator placement may therefore result in significant consequences such as excessive dose to the target, severe skin reaction, permanent discoloration, skin indentation, and poor overall cosmesis upon completion of treatment.

摘要

目的

评估在使用Xoft电子近距离放射治疗系统治疗非黑色素瘤皮肤癌过程中,施源器施加的力对百分深度剂量(PDD)和有效治疗深度的影响。

方法

为模拟可压缩组织,使用了一个5毫米厚的组织等效填充物。使用软X射线电离室进行输出测量,对所有带有塑料端盖的Xoft表面施源器进行测量。首先在施源器对填充物施加最小压力的情况下,在5毫米深度处测量输出,然后在对施源器施加均匀压力后重复测量,以计算PDD和有效治疗深度的变化。

结果

对于10毫米的锥形施源器,5牛的中等力使PDD改变超过20%。对于20毫米的锥形施源器,这种影响也很明显,而对于35毫米和50毫米的锥形施源器则最小。即使只施加中等力,有效处方深度也会改变几毫米,这与典型的处方深度相当。

结论

基于该模拟结果,施源器对皮肤施加过大压力会显著改变PDD和有效治疗深度。这种影响在10毫米和20毫米的锥形施源器中最为明显,而这两种施源器使用频率往往最高。因此,施源器放置不当可能会导致严重后果,如靶区剂量过高、严重的皮肤反应、永久性色素沉着、皮肤凹陷以及治疗结束后整体美容效果不佳。