检测骨-牙-角膜假体中的板层再吸收。

Detection of laminar resorption in osteo-odonto-keratoprostheses.

机构信息

Sussex Eye Hospital, Brighton, UK; Brighton and Sussex Medical School, Brighton, UK; Queen's Medical Centre, Nottingham, UK.

Brighton and Sussex Medical School, Brighton, UK.

出版信息

Ocul Surf. 2019 Jan;17(1):78-82. doi: 10.1016/j.jtos.2018.09.004. Epub 2018 Sep 15.

Abstract

PURPOSE

To compare the clinical examination and computerized tomography (CT) scanning methods for the detection of laminar resorption in eyes with osteo-odonto-keratoprosthesis (OOKP).

METHODS

Patients who developed laminar resorption after OOKP surgery and had at least one CT scan of the lamina during the follow-up were included. Case records and CT images and reports were retrospectively reviewed. Each lamina of the eye was regarded as a case. The imaging and clinical data were collected in Microsoft Excel, and statistical analysis was performed on Stata-v14. The agreements and sensitivities of both the methods were compared.

RESULTS

Forty patients out of 64 were found to have laminar resorption. A total of 48 laminae were studied, which had data on the presence or absence of resorption. The sensitivity and specificity of clinical detection of resorption were 87.5% (CI 68%-97%) and 37.5% (CI 8.5%-75.5%), respectively. Whereas, the sensitivity and specificity of CT scan were 81% (CI 61%-93%) and 50% (CI12%-88%), respectively. Both the methods have detected resorption in 21 out of 32 laminae having both the clinical and CT scan data. There is a fair agreement between the two techniques in the identification of thinned laminar sites.

CONCLUSIONS

Clinical detection of laminar resorption in OOKP eyes is equally sensitive to the CT scanning. Resorption can be detected even in its early stages by clinical palpation in experienced hands. Frequent CT scanning is not indicated to detect laminar resorption. Both methods complement each other. Thinned laminar segments should be compared and correlated with both the methods for full evaluation of resorption.

摘要

目的

比较临床检查和计算机断层扫描(CT)扫描在检测骨-牙-角膜-人工角膜(OOKP)术后层板吸收中的作用。

方法

纳入 OOKP 术后发生层板吸收且在随访期间至少有一次层板 CT 扫描的患者。回顾性分析病历和 CT 图像及报告。将每只眼的层板视为一个病例。将影像学和临床数据收集到 Microsoft Excel 中,并在 Stata-v14 上进行统计分析。比较两种方法的一致性和敏感性。

结果

在 64 例患者中,有 40 例发现有层板吸收。共研究了 48 个层板,这些层板均有吸收的存在或缺失的数据。临床检测吸收的敏感性和特异性分别为 87.5%(CI 68%-97%)和 37.5%(CI 8.5%-75.5%),而 CT 扫描的敏感性和特异性分别为 81%(CI 61%-93%)和 50%(CI12%-88%)。两种方法均在具有临床和 CT 扫描数据的 32 个层板中的 21 个层板中检测到吸收。两种技术在识别变薄的层板部位方面具有良好的一致性。

结论

在 OOKP 眼中,临床检测层板吸收与 CT 扫描同样敏感。在有经验的手中,即使在早期也可以通过临床触诊检测到吸收。频繁的 CT 扫描并不适用于检测层板吸收。两种方法互为补充。应该比较变薄的层板段,并与两种方法进行比较,以全面评估吸收。