Bayón J, Alejo L, Koren C, Huerga C, Corredoira E, Serrada A, Balbacid E, Gutierrez-Larraya F, Guibelalde E
Medical Physics Department, La Paz University Hospital, Madrid, Spain.
Medical Physics Department, La Paz University Hospital, Madrid, Spain.
Appl Radiat Isot. 2018 Nov;141:130-137. doi: 10.1016/j.apradiso.2018.08.014. Epub 2018 Aug 23.
This paper presents a fast method to estimate the annual eye lens dose levels for interventional practitioners applying the Monte Carlo method of uncertainty assessment. The estimation was performed by placing an anthropomorphic phantom in the typical working position, and applying the habitually employed protocol. No radiation protection devices were considered in the simulation. The results were compatible with the measurements performed during interventions by placing dosimeters in the vicinity of the eyes of two paediatric interventional cardiologists working with a fluoroscopic biplane system.
本文提出了一种快速方法,用于运用蒙特卡洛不确定性评估方法估算介入治疗从业者的年度眼晶状体剂量水平。估算过程是将一个人体模型置于典型工作位置,并应用习惯性使用的操作流程。模拟中未考虑辐射防护设备。通过在使用双平面荧光透视系统工作的两名儿科介入心脏病专家眼睛附近放置剂量计,该结果与介入治疗期间所进行的测量结果相符。