Aarhus University, Denmark.
Aarhus University Hospital, Denmark.
Health Informatics J. 2019 Dec;25(4):1815-1824. doi: 10.1177/1460458218799505. Epub 2018 Sep 25.
Preeclampsia is a serious medical disorder affecting pregnancy. Screening in early pregnancy can identify women at risk and enable effective prophylactic treatment. Accurate blood pressure (BP) measurement is an important element of the screening algorithm. Automated self-screening, while attending the first trimester ultra sound scan, using a BP self-measurement (BPSM) station, could be a low-cost alternative to office BP measurements (OBPM) on both arms performed by clinical staff, if the measurement quality can be ensured.
The aim of this study was to compare automated BPSM using a self-measurement station on one arm, with OBPM performed by clinical staff on both arms. Primary outcome was the difference in mean arterial pressure (MAP) between the two methods and secondary outcomes were safety and practicality issues.
Pregnant women attending ultrasound-examination at 12 weeks gestational age were recruited and randomized to start with having two OBPMs taken on both arms by staff, using two standard validated automatic upper arm BP devices, or self-measuring using an automated BPSM station following a crossover study design. The BPSM station consists of a validated blood pressure device, and an add-on sensor system capable of registering blood pressure values, rest-time, back-supported, legs-crossed, and ambient noise-levels respectively, and providing interactive guidance during the measurement process, for supporting the self-measurement process.
A total of 80 complete BP measurement sets were obtained, for a total of 240 BPSM measurements and 320 OBPM measurements. We found no significant difference between the OBPM and BPSM methods (p=0.86) for mean arterial pressure (MAP). However, erroneous measurements were observed frequently during the experiment, mainly during the first of the 3 BPSM measurements (6%), secondary during the second BPSM measurement (3%). Only one data set (1%) was excluded due to OBPM errors.
No significant difference in MAP between the two methods was found. Means for detecting and repeating erroneous BP measurements should be implemented. Measurement errors was found in 9 % of the measurement sets which is not acceptable for clinical use. Thus, several measures have been identified in order to properly identify and recover from such measurement errors in the future.
子痫前期是一种严重的妊娠疾病。早期妊娠筛查可以识别高危妇女,并进行有效的预防性治疗。准确的血压(BP)测量是筛查算法的重要组成部分。在进行第一次超声扫描时,使用自动自我筛查,通过 BP 自我测量(BPSM)站进行自我测量,对于临床工作人员在双臂上进行的办公室 BP 测量(OBPM)来说,可能是一种低成本的替代方法,如果可以保证测量质量的话。
本研究旨在比较使用单臂自我测量站进行的自动 BPSM 与临床工作人员在双臂上进行的 OBPM。主要结局是两种方法之间平均动脉压(MAP)的差异,次要结局是安全性和实用性问题。
招募了在 12 周妊娠时接受超声检查的孕妇,并随机分配到开始时接受临床工作人员在双臂上进行两次 OBPM,使用两个标准验证的自动上臂 BP 设备,或在交叉研究设计后使用自动 BPSM 站进行自我测量。BPSM 站由一个经过验证的血压设备和一个附加的传感器系统组成,该系统能够分别记录血压值、休息时间、背部支撑、腿部交叉和环境噪声水平,并在测量过程中提供交互式指导,以支持自我测量过程。
总共获得了 80 个完整的 BP 测量集,共计 240 个 BPSM 测量和 320 个 OBPM 测量。我们发现 OBPM 和 BPSM 方法之间的平均动脉压(MAP)没有显著差异(p=0.86)。然而,在实验过程中经常观察到错误的测量,主要是在 3 次 BPSM 测量中的第一次(6%),其次是第二次 BPSM 测量(3%)。只有一个数据集(1%)因 OBPM 错误而被排除。
两种方法之间 MAP 没有显著差异。应该实施检测和重复错误 BP 测量的方法。在 9%的测量集中发现了测量误差,这对于临床使用来说是不可接受的。因此,已经确定了一些措施,以便在未来正确识别和从这种测量误差中恢复。