采用快速千伏切换双能计算机断层扫描对肾结石成分进行特征描述与实验室结石分析的对比:一项初步研究。

Characterization of renal stone composition by using fast kilovoltage switching dual-energy computed tomography compared to laboratory stone analysis: a pilot study.

机构信息

Division of Urology, Department of Surgery, Faculty of Medicine, Siriraj Hospital, Mahidol University, 2 Wanglang Road, Bangkoknoi, Bangkok, 10700, Thailand.

Department of Radiology, Faculty of Medicine, Siriraj Hospital, Mahidol University, Bangkok, Thailand.

出版信息

Abdom Radiol (NY). 2019 Mar;44(3):1027-1032. doi: 10.1007/s00261-018-1787-6.

Abstract

PURPOSE

To prospectively examine the diagnostic performance of fast kilovoltage switching dual-energy computed tomography (DECT) in characterization of in vivo renal stone composition compared with postoperative stone analysis.

METHODS

Consecutive consenting patients scheduled for endoscopic kidney stone surgery in a tertiary referral hospital from June 2015 to January 2016 were enrolled. Patients were preoperatively scanned with single-source, fast kilovoltage switching DECT. Stone compositions were determined regarding the effective atomic number measurements. Results of the stone compositions from DECT were compared to postoperative infrared spectroscopy stone analysis as the standard reference.

RESULTS

For the 39 patients enrolled in the study, DECT was able to detect uric acid stone with sensitivity of 88.9% and specificity of 100%. There was 100% positive predictive value, 96.8% negative predictive value, and 97.4% accuracy. For non-contrast CT scan, sensitivity was 88.9%, specificity was 96.7%, positive predictive value was 88.9%, negative predictive value was 96.7%, and accuracy was 94.8%. Of the 39 samples examined, 21 (54%) were single composition, whereas 18 (46%) were combined. Single composition stones were correctly characterized by DECT in 100% (8/8) for uric acid. Whereas the result of uric acid stone containing stone discrimination in mixed composition was not so good with Z alone, iodine imaging can compensated this fault.

CONCLUSIONS

DECT provides excellent accuracy in characterizing uric acid stone compositions. With the addition of iodine image, all of uric acid-containing stones can be determined by the DECT.

摘要

目的

前瞻性研究快速千伏切换能谱 CT(DECT)在体内肾结石成分特征描述中的诊断性能,并与术后结石分析进行比较。

方法

连续纳入 2015 年 6 月至 2016 年 1 月期间在一家三级转诊医院接受内镜肾结石手术的患者。所有患者术前均行单源快速千伏切换 DECT 扫描。根据有效原子数测量结果确定结石成分。将 DECT 得到的结石成分结果与术后红外光谱结石分析作为标准参考进行比较。

结果

在这项研究中,共纳入 39 例患者,DECT 检测尿酸结石的敏感度为 88.9%,特异度为 100%。其阳性预测值为 100%,阴性预测值为 96.8%,准确率为 97.4%。非增强 CT 扫描的敏感度为 88.9%,特异度为 96.7%,阳性预测值为 88.9%,阴性预测值为 96.7%,准确率为 94.8%。在检查的 39 个样本中,21 个(54%)为单一成分,18 个(46%)为混合成分。DECT 正确识别了 100%(8/8)的尿酸结石的单一成分,而对于混合成分中尿酸结石的鉴别结果则不理想,Z 值成像不能很好地反映。碘成像可以弥补这一缺陷。

结论

DECT 在描述尿酸结石成分方面具有出色的准确性。通过添加碘图像,可以通过 DECT 确定所有含尿酸的结石。