Blue Matt, Douthit Christian, Dennison Joel, Caroom Cyrus, Jenkins Mark
Department of Orthopaedic Surgery and Rehabilitation, Texas Tech University Health Sciences Center, 3601 4th St. MS 9436, Lubbock, TX 79436, USA.
Case Rep Orthop. 2018 Sep 16;2018:2381406. doi: 10.1155/2018/2381406. eCollection 2018.
A rare complication from computer-navigated total knee arthroplasty is a fracture through the insertion site of a tracking pin. These pins are inserted across the femoral and tibial shafts either bicortically, transcortically, or unicortically and have a reported fracture incidence of 1.38%, with all published cases occurring after bicortical pin placement. In this case, a 60-year-old female suffered a femoral shaft fracture through a unicortically inserted computer navigation tracking pin 6 weeks after total knee arthroplasty. Her fracture was successfully fixated with an intramedullary nail with retention of the knee prosthesis. This case is important as it records the risk for a postoperative fracture through a unicortically inserted computer navigation pin.