Suppr超能文献

儿童患者中 Rhythmia 标测系统的安全性和准确性。

Safety and accuracy of the Rhythmia mapping system in pediatric patients.

机构信息

Department of Pediatric Cardiology, Saitama Medical University International Medical Center, Saitama, Japan.

Department of Pediatric Cardiology, Saitama Medical University International Medical Center, Saitama, Japan.

出版信息

Heart Rhythm. 2019 Mar;16(3):388-394. doi: 10.1016/j.hrthm.2018.10.002. Epub 2018 Oct 9.

Abstract

BACKGROUND

A new mapping system (Rhythmia) using a 64 mini-electrode small basket array (Orion) was developed that enables rapid high-density mapping in a short time. However, there are few reports about the usefulness of this system in pediatric cases.

OBJECTIVE

The purpose of this study was to investigate the safety and accuracy of the Rhythmia system and Orion catheter in children.

METHODS

Catheter ablation was performed using the Rhythmia system and Orion catheter in 23 patients younger than 20 years (body weight >20 kg) without a past medical history of cardiac disease. Mapping time, number of mapping beats, and number of mapping electrodes were compared for left atrium, right atrium and right ventricular outflow tract.

RESULTS

Twenty-three maps of the right atrium were acquired in 12.6 minutes (range 8.9-15.1), consisting of 709 beats (range 492-1163) and 7132 electrograms (range 4618-10,533). Twelve maps of the left atrium were acquired in 12.1 minutes (range 9.8-14.6), consisting of 565 beats (range 446-881) and 6412 electrograms (range 4912-11,402). There were no significant difference in mapping time, accepted beats, and electrograms between the 2 chambers. Manual annotation was needed in 53 of 293,185 electrograms (0.01%) due to far-field ventricular electrogram sensing and artifact. No adverse events occurred in any of the cases.

CONCLUSION

The Orion catheter and Rhythmia mapping system were safe and accurate for mapping various arrhythmias in pediatric patients. Detailed geometry and high-resolution activation mapping were acquired without the need for manual reannotation.

摘要

背景

开发了一种新的映射系统(Rhythmia),该系统使用 64 个迷你电极小篮阵列(Orion),可在短时间内实现快速高密度映射。然而,关于该系统在儿科病例中的实用性的报告很少。

目的

本研究旨在探讨 Rhythmia 系统和 Orion 导管在儿童中的安全性和准确性。

方法

对 23 例无心脏病既往史的体重>20kg 的 20 岁以下患者使用 Rhythmia 系统和 Orion 导管进行导管消融。比较左心房、右心房和右心室流出道的映射时间、映射节拍数和映射电极数。

结果

在 12.6 分钟(8.9-15.1 分钟)内获得了 23 个右心房图,包含 709 个节拍(492-1163 个)和 7132 个电图(4618-10533 个)。在 12.1 分钟(9.8-14.6 分钟)内获得了 12 个左心房图,包含 565 个节拍(446-881 个)和 6412 个电图(4912-11402 个)。两个腔室之间的映射时间、接受节拍和电图没有显著差异。由于远场心室电图感应和伪影,需要对 293185 个电图中的 53 个(0.01%)进行手动注释。在任何情况下均未发生不良事件。

结论

Orion 导管和 Rhythmia 映射系统在儿科患者中用于各种心律失常的映射是安全且准确的。无需手动重新注释即可获得详细的几何形状和高分辨率激活映射。