PLoS One. 2018 Oct 25;13(10):e0206304. doi: 10.1371/journal.pone.0206304. eCollection 2018.
To assess the concurrent validity of a wireless patch sensor to monitor time lying, sitting/standing, and walking in an experimental and a hospital setup.
Healthy adults participated in two testing sessions: an experimental and real-world hospital setup. Data on time lying, sitting/standing, and walking was collected with the HealthPatch and concurrent video recordings. Validity was assessed in three ways: 1. test for mean differences between HealthPatch data and reference values; 2. Intraclass Correlation Coefficient analysis (ICC 3.1 agreement); and 3. test for mean differences between posture detection accuracies.
Thirty-one males were included. Significant mean differences were found between HealthPatch data and reference values for sitting/standing (mean 14.4 minutes, reference: 12.0 minutes, p<0.01) and walking (mean 6.4 minutes, reference: 9.0 minutes, p<0.01) in the experimental setup. Good correlations were found between the HealthPatch data and video data for lying (ICC: 0.824) and sitting/standing (ICC: 0.715) in the hospital setup. Posture detection accuracies of the HealthPatch were significantly higher for lying and sitting/standing in the experimental setup.
Overall, the results show a good validity of the HealthPatch to monitor lying and poor validity to monitor sitting/standing or walking. In addition, the validity outcomes were less favourable in the hospital setup.
评估无线贴片传感器在实验和医院环境中监测卧床、坐/站和行走时间的同时效度。
健康成年人参与了两个测试环节:实验和真实医院环境。使用 HealthPatch 和同期视频记录来收集卧床、坐/站和行走时间的数据。通过三种方法评估效度:1. 测试 HealthPatch 数据与参考值之间的均值差异;2. 组内相关系数分析(ICC 3.1 一致性);3. 测试姿势检测准确率之间的均值差异。
共纳入 31 名男性。在实验环境中,HealthPatch 数据与参考值在坐/站(平均 14.4 分钟,参考值 12.0 分钟,p<0.01)和行走(平均 6.4 分钟,参考值 9.0 分钟,p<0.01)方面存在显著的均值差异。在医院环境中,HealthPatch 数据与视频数据在卧床(ICC:0.824)和坐/站(ICC:0.715)方面存在良好的相关性。在实验环境中,HealthPatch 对卧床和坐/站的姿势检测准确率显著更高。
总体而言,研究结果表明 HealthPatch 监测卧床的有效性较好,而监测坐/站或行走的有效性较差。此外,在医院环境中,效度结果不太理想。