The aim of this study was to determine the impact of adaptive iterative dose reduction 3 D (AIDR3D) on the improvement of shoulder image quality in low-radiographic dose head and neck CT angiography (CTA).
Ninety patients who underwent CTA examination were randomly divided into two groups, namely group A (n = 45) and B (n = 45). Patients in group A were scanned under 120 kV and 300 mA, with images reconstructed by filtered back projection (FBP), and patients in group B were scanned under 80 kV and auto mA with images reconstructed by AIDR3D. Image quality was accessed by two experienced radiologists. The noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of common carotid artery (CCA) at C7 level, and radiation dosage were compared between the two groups.
The score of CCA in group B was significantly higher than group A (p < 0.05), and there were no significant differences in the scores of carotid sinus and internal carotid artery between the two groups (p > 0.05). The score of intracranial artery in group B was lower than that of group A, however, the image quality in group B can meet the requirement of clinical diagnosis. The noise value of CCA at C7 level in group B was significantly lower than that of group A (p < 0.05). SNR and CNR values of CCA at C7 level in group B were significantly higher than those of group A (p < 0.05). Effective radiation dose in group B was significantly decreased compared with group A (p < 0.05).
AIDR3D remarkably improved image quality in low-radiographic dose head and neck CTA over FBP, which made the low-dose CTA images meet the requirement of clinical diagnosis.
本研究旨在探讨自适应迭代剂量降低 3D(AIDR3D)在降低头颈部 CT 血管造影(CTA)低剂量辐射剂量的同时,对改善肩部图像质量的影响。
90 例行 CTA 检查的患者随机分为两组,A 组(n=45)和 B 组(n=45)。A 组患者采用 120kV 和 300mA 扫描,滤波反投影(FBP)重建图像;B 组患者采用 80kV 和自动 mA 扫描,AIDR3D 重建图像。由两位有经验的放射科医生评估图像质量。比较两组患者颈总动脉(CCA)于 C7 水平的噪声、信噪比(SNR)和对比噪声比(CNR)以及辐射剂量。
B 组 CCA 评分明显高于 A 组(p<0.05),而两组颈动脉窦和颈内动脉评分无明显差异(p>0.05)。B 组颅内动脉评分低于 A 组,但 B 组图像质量能满足临床诊断要求。B 组 CCA 于 C7 水平的噪声值明显低于 A 组(p<0.05)。B 组 CCA 于 C7 水平的 SNR 和 CNR 值明显高于 A 组(p<0.05)。B 组有效辐射剂量明显低于 A 组(p<0.05)。
与 FBP 相比,AIDR3D 可显著提高低剂量头颈部 CTA 的图像质量,使低剂量 CTA 图像满足临床诊断要求。