Mishra A C, Parks N J
Int J Rad Appl Instrum A. 1987;38(6):455-61. doi: 10.1016/0883-2889(87)90099-2.
We have demonstrated that commercially available liquid scintillation counters are adequate for obtaining useful information about the variation of beta spectral shapes as a function of attenuation. A gel scintillation system that permitted the approximation of point source geometry worked well. The spectral shapes predicted by the Fermi equations were reasonably matched with experimental spectra of sources with geometry close to a point source. For energies in the range of 50-2300 keV, we found sufficient linearity of the energy scale and uniformity of detection limits for the direct output of the instrument to be experimentally useful. The gel scintillation system has been used to directly measure the energy distribution of the beta flux coming out of bone samples and incident on the soft tissue of dogs fed 90Sr-90Y from in utero to 540 days of age.
我们已经证明,市售的液体闪烁计数器足以获取有关β谱形状随衰减变化的有用信息。一种允许近似点源几何形状的凝胶闪烁系统运行良好。费米方程预测的谱形状与几何形状接近点源的源的实验光谱合理匹配。对于50 - 2300 keV范围内的能量,我们发现仪器直接输出的能量刻度具有足够的线性和检测限的均匀性,在实验上是有用的。凝胶闪烁系统已被用于直接测量从子宫内到540日龄喂食90Sr - 90Y的狗的骨样品中射出并入射到软组织上的β通量的能量分布。