Sensors (Basel). 2018 Nov 1;18(11):3721. doi: 10.3390/s18113721.
Texas Instruments ADS1299 is an attractive choice for low cost electroencephalography (EEG) devices owing to its low power consumption and low input referred noise. To date, there have been no rigorous evaluations of its performance. In this EEG experimental study we evaluated the performance of the ADS1299 against a high quality laboratory-based system. Two self-paced lower limb motor tasks were performed by 22 healthy participants. Recorded power across delta, theta, alpha, and beta EEG bands, the power ratio across the motor tasks, pre-movement noise, and signal-to-noise ratio were obtained for evaluation. The amplitude and time of the negative peak in the movement-related cortical potentials (MRCPs) extracted from the EEG data were also obtained. Using linear mixed models, no statistically significant differences ( > 0.05) were found in any of these measures across the two systems. These findings were further supported by evaluation of cosine similarity, waveform differences, and topographic maps. There were statistically significant differences in MRCPs across the motor tasks in both systems. We conclude that the performance of the ADS1299 in combination with wet Ag/AgCl electrodes is analogous to that of a laboratory-based system in a low frequency (<40 Hz) EEG recording.
德州仪器 ADS1299 由于其低功耗和低输入参考噪声,是低成本脑电图 (EEG) 设备的理想选择。迄今为止,尚未对其性能进行严格评估。在这项 EEG 实验研究中,我们评估了 ADS1299 与高质量实验室系统的性能。22 名健康参与者执行了两个自主下肢运动任务。记录了 delta、theta、alpha 和 beta EEG 频段的功率、运动任务之间的功率比、运动前噪声和信噪比,以进行评估。还从 EEG 数据中提取了运动相关皮质电位 (MRCP) 的负峰的幅度和时间。使用线性混合模型,在这两个系统中,这些指标的任何一个都没有发现统计学上显著的差异(>0.05)。余弦相似性、波形差异和地形图的评估进一步支持了这些发现。在两个系统中,运动任务之间的 MRCP 存在统计学上的显著差异。我们得出结论,ADS1299 与湿 Ag/AgCl 电极结合的性能类似于低频 (<40 Hz) EEG 记录的实验室系统。