Abstract

INTRODUCTION

The purpose of this study was to assess temperature development in endodontic sealers during different obturation techniques in a closed system simulating the surrounding biological structures at body temperature.

METHODS

The root canals of 48 human single-rooted maxillary canines were instrumented with ProTaper Gold (Dentsply Sirona, York, PA) to size F3. On the buccal aspect of each root, artificial side canals with a diameter of 0.5 mm were drilled at a distance of 3, 6, and 9 mm from the apical foramen, and type K thermocouples were inserted via plastic pipes. The roots were positioned in plastic vials filled with alginate. The root canals were obturated by the continuous wave and warm backfill technique, Thermafil obturators (Dentsply Sirona), or single-cone obturation (n = 12) at body temperature using AH Plus sealer (Dentsply Sirona). Temperature measurement during the obturation procedure was assessed by thermocouples. Statistical analysis of the maximum temperature change was performed using the Kruskal-Wallis test (P = .05).

RESULTS

The continuous wave and warm backfill technique caused significantly higher temperatures than Thermafil and single-cone obturation (P < .05). The continuous wave technique revealed significantly higher temperatures than the warm backfill technique at 3 mm and 6 mm from the apex (P < .05).

CONCLUSIONS

In a closed system with simulated surrounding tissues at 37°C, the continuous wave technique produced higher increases in temperature than the warm backfill technique with a maximum of 19.1°C. The temperature increase during Thermafil obturation was negligible. The temperature increase in the endodontic sealer was markedly lower than expected from root surface temperature measurement studies.

摘要

简介

本研究旨在评估在模拟体温周围生物结构的封闭系统中,不同根管充填技术下牙胶封闭剂的温度变化。

方法

48 颗上颌单根恒磨牙的根管用 ProTaper Gold(登士柏西诺德,约克,宾夕法尼亚州)锉至 F3 号。在每个牙根的颊侧,距根尖孔 3、6 和 9mm 处用直径为 0.5mm 的人造侧支钻,并用塑料管插入 K 型热电偶。将牙根置于充满藻酸盐的塑料小瓶中。在体温下,用连续波和热回填技术、Thermafil 热牙胶充填器(登士柏西诺德)或单根管充填技术(n=12)对根管进行充填,使用 AH Plus 封闭剂(登士柏西诺德)。通过热电偶评估根管充填过程中的温度测量。采用 Kruskal-Wallis 检验(P=.05)对最大温度变化进行统计学分析。

结果

连续波和热回填技术引起的温度明显高于 Thermafil 和单根管充填(P<.05)。连续波技术在距根尖 3mm 和 6mm 处的温度明显高于热回填技术(P<.05)。

结论

在 37°C 的模拟周围组织的封闭系统中,连续波技术比热回填技术引起的温度升高更高,最高可达 19.1°C。Thermafil 热牙胶充填的温度升高可忽略不计。牙胶封闭剂的温度升高明显低于从根管表面温度测量研究中预期的温度升高。