瑞姆通气鼻塞管在早产儿中提供持续气道正压通气的效果是否与 Hudson 鼻罩一样有效?

Does the RAM Cannula Provide Continuous Positive Airway Pressure as Effectively as the Hudson Prongs in Preterm Neonates?

机构信息

Division of Neonatology, Department of Pediatrics, Dartmouth Hitchcock Medical Center, Lebanon, New Hampshire.

Department of Respiratory Therapy, Dartmouth Hitchcock Medical Center, Lebanon, New Hampshire.

出版信息

Am J Perinatol. 2019 Jul;36(8):849-854. doi: 10.1055/s-0038-1675330. Epub 2018 Nov 5.

Abstract

OBJECTIVE

To compare the level of continuous positive airway pressure (CPAP) delivered by the RAM cannula system (Neotech, Valencia, CA) with that delivered by a traditional CPAP nasal delivery interface (Hudson prongs; Hudson-RCI, Temecula, CA) in preterm infants with respiratory distress.

METHODS

This was a crossover intervention study in a convenience sample of preterm infants with respiratory distress requiring treatment with CPAP. We measured the mean intraoral (pharyngeal) pressure, which approximates the applied CPAP level, using both the RAM cannula and Hudson prongs. The primary outcome was a comparison of the differences between the set CPAP levels and the measured intraoral pressures of both delivery systems.

RESULTS

We analyzed data from 12 preterm infants with mean (standard deviation) birth weight of 1,225 (405) g and gestational age of 28.4 (2.1) weeks at a median postnatal age of 10 days. The mean difference (95% confidence interval) between the set CPAP level and measured intraoral pressure was -2.45 cm HO (-3.36, -1.55) with the RAM cannula and +0.40 cm HO (-0.30, 1.12) with Hudson prongs,  = 0.0002.

CONCLUSION

For given set CPAP pressure level in preterm infants, the RAM cannula system consistently delivers lower pharyngeal pressure (effective CPAP) levels than Hudson prongs.

摘要

目的

比较经 RAM 管系统(Neotech,加利福尼亚州瓦伦西亚)输送的持续气道正压(CPAP)与传统 CPAP 鼻输送接口(Hudson 叉;Hudson-RCI,加利福尼亚州特美库拉)输送的 CPAP 在有呼吸窘迫的早产儿中的水平。

方法

这是一项在有呼吸窘迫需要 CPAP 治疗的早产儿便利样本中进行的交叉干预研究。我们使用 RAM 管和 Hudson 叉测量了接近应用 CPAP 水平的口腔内(咽)压力平均值。主要结果是比较两种输送系统的设定 CPAP 水平和测量的口腔内压力之间的差异。

结果

我们分析了 12 名早产儿的数据,他们的平均(标准差)出生体重为 1225(405)g,胎龄为 28.4(2.1)周,中位数出生后年龄为 10 天。RAM 管的设定 CPAP 水平与测量的口腔内压力之间的平均差值(95%置信区间)为-2.45cmHO(-3.36,-1.55),而 Hudson 叉为+0.40cmHO(-0.30,1.12), = 0.0002。

结论

在早产儿中,对于给定的设定 CPAP 压力水平,RAM 管系统输送的咽压力(有效 CPAP)水平始终低于 Hudson 叉。